A.1.2 tob1 – Campbell Scientific RTDAQ Software User Manual

Page 358

Advertising
background image

Appendix A. Campbell Scientific File Formats

An example showing a TOA5 file containing the optional timestamp and
record number:

"TOA5","CR1000","CR1000","1031","CR1000.Std.00.60","CPU:Test.CR1","4062","Test"

"TIMESTAMP","RECORD","batt_volt_Min","PTemp"

"TS","RN","Volts","C"

"","","Min","Smp"

"2004-11-11 15:03:45",0,13.7,24.92

"2004-11-11 15:04:00",1,13.7,24.95

"2004-11-11 15:04:15",2,13.7,24.98

A.1.2 TOB1

TOB1 files can be generated by RTDAQ when saving data files to the PC.
This binary file format is typically only used when it is essential to minimize
the file size or when other software requires this format. It has the following
structure:

ASCII header line 1

ASCII header line2

ASCII header line3

ASCII header line 4

ASCII header line5

Binary Records

An example of a TOB1 ASCII header:

"TOB1","STATION","CR9000","1000","1.00","CPU:BIG.DLD","25871","VALUES"

"SECONDS","NANOSECONDS","RECORD","Array(1)","Array(2)","Fast","my_string"

"","","RN","mVolts","mVolts","mVolts"

"","","","Smp","Smp","Smp"

"ULONG","ULONG","ULONG","IEEE4","IEEE4","FP2","ASCII(25)"

Header line one describes the file environment with the following eight fields:

• Data file type (TOB1).

• Station name (STATION).

• Model name of the datalogger (CR9000).

• Serial number of the datalogger (1000).

• Operating system on the datalogger (1.00).

• Name of the program running in the datalogger (CPU:BIB.DLD).

• Signature of the program running in the datalogger (25871).

• The name of the datalogger table (VALUES).

A-2