LINEARITY

Electronics: ±0.05% (±0.01pH units / ±1mV of full scale)

RESOLUTION

Digital Display: ±0.01 pH units / ±1mV

Analog Meter: ±0.13pH units / ±40mV

SENSITIVITY

0.05 % of full scale (0.01pH units / 1mV)

STABILITY

±0.05 % of full scale (±0.01pH units / ±1mV)

REPEATABILITY

±0.05 % of full scale (±0.01pH units / ±1mV)

CALIBRATION CHECK

Built in (7 & 0 pH / 0 & -2000mV)

RECORDER OUTPUT

Isolated 0-10 VDC @ 5mA max. (linear); standard on all

models

OUTPUT IMPEDANCE

100Ω ±5%

SENSOR INPUT

1 (optional 3 sensor Input available)

NOTE: pH and ORP may NOT be mixed.

CONTROL FUNCTIONS

Models 722II, 723II & 727II, 728II:

Single set point alarm/control continuously adjustable

0-100% of span

Second set point alarm/control (Optional).

Continuously adjustable - 0-100% of span

Hysteresis

Adjustable from 0.3-3 % of full scale

Indicators

“HIGH” (red) and “LOW” (green) set point LEDs - reversible.

Relay Contact Rating

SPDT 10 amp @ 250 VAC, 30 VDC. Relay operates

increasing or decreasing reading (user selectable).

Solid State Output

24VDC unregulated, 30mA Maximum.

Powers optional PA - Piezo Electric Alarm, RA™ - Remote

Alarm, or customer alarm within above limits.

Second Alarm/control Module, with above specifications, opt. on

Models 722II, 723II & 727II, 728II:

POWER SPECIFICATIONS

115/230 VAC ±15%, 50/60 Hz (User selectable)

overvoltage category II

100 mA Maximum Current

Double Insulated (with circuit board ground for OEM operation)

Fuse - 100mA Slow Blow (T.10A) for both 115 & 230 VAC (V~)

User replaceable

Humidity - 20-90% non-condensing

Max. Altitude -

40,000 ft/12,000 meters non-operating

10,000 ft/3000 meters operating

Pollution degree 2

24 VAC or 24 VDC Option available on Special Order

Overvoltage category II (24 VAC)

250 mA Maximum Current

Double Insulated (with circuit board ground for OEM operation)

Fuse - 250mA Slow Blow (T.25A) User replaceable

AMBIENT TEMPERATURE RANGE

32°F to 140°F ( 0°C to 60°C)

DIMENSIONS

6.0” (152mm) H x 4.8” (122mm) W x 3.8” (96mm) D

HOUSING CONSTRUCTION

Fully gasketed heavy-duty ABS for splashproof and corrosion

resistance.

Rated IP65/NEMA 4X

Double Insulated

WEIGHT

720 Series II: average 2 Ibs. (0.9 kg)

C. OPTIONAL FEATURES

-SC

Second Alarm/control Module (M/c models only)

-4A

4-20 mA Isolated output (Digital & OEM only)

-PA

70 db Piezo Electric Alarm (digital M/c models only)

-PAT

Piezo Alarm & Timer Module (digital M/c models only)

-35BL 3 1/2 digit backlit LCD

-TP

Temperature Module Requires -TP sensor

(Digital Models Only)

-TH

Alarm/control Interface Harness (requires SC & TP)

-PC

115 VAC Powercord (8 ft. with USA plug and strain

relief) - NOT for use with 230 VAC.

-24VA 24 VAC isolated power supply (special order)

-24VD 24 VDC isolated power supply (special order)

-PTST Panel mounted ZERO (7pH or 0mV) Test switch

(special order)

D. ACCESSORIES (ordered separately)

SCO

Second Alarm/control Module (M/c models only)

4AO

4-20 mA Isolated output (Digital & OEM only)

3SO

3 Sensor Input Switch Module with

3 foot/1 meter cable*.

3SE

3 Sensor Input Switch with enclosure.

PAO

70 db Piezo Electric Alarm only (723II/728II)

PATO Piezo Alarm & Timer Module (723II/728II)

35BLO 3 1/2 digit backlit LCD w/Bezel

TPO

Temperature Module Requires additional ITS1

sensor (Digital Models Only)

THO

Alarm/control Interface Harness

(requires SCO & TPO)

PCO

115 VAC Powercord (8 ft. with USA plug and strain

relief) - NOT for use with 230 VAC.

SMP50 Surface Mounting Plate

RA

Remote Alarm - RA™ (controller models only)

VR

Powersupply, 24 VAC, 20VA (115 VAC, indoor use)

024-1 24 VAC 1” Solenoid Valve

720SS pH/ORP Sensor Substitute (NIST Traceable)

pHC

NIST Certificate without Sensor

pHCS NIST Certificate with Sensor in Buffer Solutions

OC

NIST Certificate without Sensor

*Customer mounted separately

6

DOUBLE INSULATED