G. RE-CONFIGURE YOUR MONITOR/CONTROLLER

(Configuration Module Installation)

1. DESCRIPTION

The 720 Series II Monitor/controllers have been designed for

easy field re-configurability. The Configuration Module consists of

a 16 pin Header that plugs into a 16 pin socket.

NOTE: Changing from pH to ORP or visa versa, requires a

different sensor.

pH Module

pH14

ORP Module

ORP2A ANALOG Module†

ORP Module

ORP2D DIGITAL Module†

NOTE: Some OEM models may not be re-configured being

originally manufactured with fixed resistors. If your application

requires re-configuring an OEM model, first contact the system

manufacture for help. The Myron L Company may re-configure or

exchange your instrument at a cost.

2. INSTALLATION

Briefly -

The new Configuration Module simply replaces the Configuration

Module presently installed, see figure II.G.1.

Type label will be required if changing from pH to ORP or visa

versa.

ZERO and SPAN recalibration is performed.

IMPORTANT NOTES:

1. CHANGING FROM pH TO ORP OR VISA VERSA

WILL REQUIRE A SCALE CHANGE ON ANALOG

MODELS. SEE SECTION II.G.4 FOR INSTRUCTIONS.

† 2. ANALOG and DIGITAL ORP MODELS REQUIRE

DIFFERENT CONFIGURATION MODULES. SEE

FIGURES II.G.8 AND II.G.9.

WARNING: BEFORE STARTING, IF MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER

IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE

TO THE INSTRUMENT, AND COULD BE HARMFUL OR

FATAL

TO

PERSONNEL.

ONLY

QUALIFIED

PERSONNEL

SHOULD

INSTALL

OR

SERVICE

ELECTRICAL EQUIPMENT.

Physical

NOTE: When opening instrument, remove front cover with care;

a ribbon cable connects the front panel and main board. If the

front panel has all ready been removed from the enclosure skip to

#4.

1. Using a standard slot screwdriver remove the

screws on the front panel.

2. Carefully wiggle the front panel to loosen the gasket and

pull gently toward you. Do not pull more than about 8

inches/20CM or you could damage the wiring harness.

3. Turn the front panel around so that the back side is

facing you and set aside.

4. Locate and remove existing Configuration Module from

MAIN Circuit Board, as shown in figure II.G.1. It is not

easy to remove, it was designed to stay in place under

adverse conditions.

5. With the pointer up, carefully align the new configuration

Module to the socket on the MAIN Circuit Board as shown

in figure II.G.1.

6. Press firmly into place.

7. Analog models will require the meter scale to be changed.

See Changing Analog Meter Scale, section II.G.3.

8. Recalibrate, see CALIBRATION PROCEDURES, V.C.

16

!

CAUTION - READ FOLLOWING CAREFULLY

Figure II.G.1

REMOVE TO

INSTALL

FUSE*

115/

230

TRANSFORMER

Main CB Assembly

721 726

722 727

723 728

724 729

MYRON L

COMPANY

CONFIGURATION

MODULE

UP

}

3S

INC

DEC

SPC

SPAN

ZERO

HYS1

SP1

SPAN

ZERO

POINTER

DO NOT

ADJUST

0-10VDC

OUTPUT

7 pH/ZERO mV

PUSH TO TEST

DIS

ORP ZERO

SPAN (GAIN )

PUSH TO TEST

BK WT RD GN NU R- R+

PWR

C GD NC NO CM

PA

RA

GN

D

CH

S

REMOVE T

O INST

ALL

SECOND RELA

Y

pH

}

SP

AN

DE

C

Advertising
This manual is related to the following products: