3. CHANGING ANALOG METER SCALE (DIAL)

After changing configurations (pH to ORP or visa versa) on an

analog model 720II it will be necessary to change the meter scale

(dial). The following steps will guide you through the change.

Please read caution carefully.

Exercise extreme caution while working on meter.

Meter movement, and/or pointer (needle) may be

damaged beyond repair. It is recommended only

qualified personnel change scales.

DO NOT allow dust to enter the case while it is open.

Failure to follow these instructions may cause

irreparable damage, and will VOID the warranty.

Briefly -

The meter is removed from front panel (except OEM models).

The cover is removed.

The hex screws retaining the scale are removed, and scale

is slid out the top.

New scale is slid in place, and screws reinstalled.

Cover aligned, installed and zero set.

Physical

For OEM models skip to #4.

1. Using a wrench, nut driver or pliers remove the two (2)

large nuts retaining the circuit board wiring harness

connectors. Red wire is PLUS (+),

Brown wire is MINUS (-). See figure II.G.2.

2. Using a standard slot screwdriver remove the four (4)

screws, and the backing plate holding the meter to the

front panel as shown in figure II.G.2.

3. Remove meter. Do not lose the gasket between the

meter and the front panel, it must be reinstalled upon

reassembly. See figure II.G.3. Align gasket to front

panel so as the gasket does not show when meter is

assembled to front panel.

4. Using a wide blade standard slot screwdriver, starting at

a bottom corner, CAREFULLY pry the clear plastic cover

loose from the movement case. See figure II.G.4.

5. Using a 1/8” (3.2mm) hex driver, remove the two (2) small

screws holding the scale (dial). See figure II.G.5.

6. To remove the scale, lightly press on bottom of scale,

the top will lift up. Be extremely careful not to bend the

pointer (needle) or to move the Zero Adjust (ZA).

See figure II.G.5.

7. Grasp top and slide scale slowly out from top. Be

extremely careful not to bend the pointer (needle) or to

move the Zero Adjust (ZA). See figure II.G.5.

17

CLEAR COVER

Pry Here Carefully

CASE

Figure II.G.4

Figure II.G.5

Grasp Scale Here

and Slide Out TOP

SCALE (dial)

Pointer (needle)

Inverted "Y"

1/8" (3.18mm)

Hex Screws (2)

METER CASE

Press

Lightly

Here

0

!

CAUTION - READ FOLLOWING CAREFULLY

GASKET

Figure II.G.3

Figure II.G.2

Harness to Main

CB Connector

+

Backing Plate

METER

Mounting Screws (4)

BROWN WIRE

RED WIRE