Panel mounted display, Display, Temperature display – Myron L 729II User Manual

Page 37

Advertising
background image

6. OEM INSTALLATION

b. Dual Display Models SPECIAL ORDER

These models allow simultaneous display of both Cond/TDS, RES

or pH/ORP and Temperature. They require two display harnesses.

The Cond/TDS, RES or pH/ORP display harness is part number

DH2DD without control feature, or part number DHDD with control

feature. The Temperature display harness is part number TPMH.

The DH2DD or DHDD is a “Y” harness that is connected from the

main CB to the Cond/TDS, RES or pH/ORP display while the other

end is connected to the Temperature Module as shown in figure

III.C.15.

The Temperature Module display harness TPMH connects to a

separate Temperature display, the same as in figure III.C.15.

33

MYRON L

COMPANY

TP/TPO - TEMPERATURE MODULE

7 pH/ZERO mV

PUSH TO TEST

SET POINT

#1 CONTROL

SET POINT #1

HYSTERESIS

RIGHT INC

LEFT DEC

SPAN (GAIN)

PUSH TO TEST

PANEL MOUNTED DISPLAY

REAR VIEW

Figure III.C.15

CHS

GND

UP

}

FUSE*

115/

230

PWR

C GD NC NO CM

BK WT RD GN NU R- R+

3S

PA

RA

DEC

SPC

Main CB Assembly

0-10VDC

OUTPUT

TRANSFORMER

FS

ZERO

DIS

SP2

HYS2

DISPLAY CONNECTION

DISPLAY

SELECT

INC

INC

DEC DEC

TEMP

CONTROL

HARNESS

CONNECT

MAIN

CIRCUIT

BOARD

DISPLAY

HARNESS

HERE

JUMPER

CONNECT

FRONT

PANEL

DISPLAY

HARNESS

(TPMH)

HERE

TEMP

SENSOR

(-)

(+)

BRN

BLU

BR BL OR

-

+

+

0-5VDC

-

OR

COND/TDS , RES or pH/ORP

DISPLAY

DISPLAY CONNECTION

TEMPERATURE

DISPLAY

REMOVE T

O INST

ALL

SECOND RELA

Y

SPC

pH

CONFIGURATION

MODULE

721 726

722 727

723 728

724 729

DO NOT

ADJUST

HYS1

SP1

SP

AN

ZERO

ORP ZERO

DIS

INC

ZERO SPAN

DISPLAY CONNECTION

Advertising
This manual is related to the following products: