Myron L 729II User Manual

Page 53

Advertising
background image

49

V. COMPONENT IDENTIFICATION,

CALIBRATION AND PREVENTIVE CARE

B.

METER MECHANICAL ZERO PROCEDURES

ANALOG METERS ONLY

1. Ensure power is OFF.

2. Note the position of the meter needle. If it has come to

rest pointing at the zero mark on the scale, turn the

Monitor back ON and continue on to Section V.C.

3. Locate the small (approx. 1/4”), black plastic button

directly below the center of the meter. Use a small

screwdriver or a fingernail to gently pry it loose and

remove it from the access hole. See figure II.G.9.

4. Insert a small standard screwdriver into the access hole

and carefully locate the slot in the mechanical

adjustment plug.

5. Turn the adjustment plug slightly until the needle rests

on zero.

6. Insert the small plastic button into the access hole.

7. Turn power ON.

C.

CALIBRATION PROCEDURES

All Myron L pH/ORP Monitor/controllers are factory calibrated

prior to shipment and are ready to install without further

calibration. Calibration should be checked occasionally with the

internal ZERO Test switch (ZERO) to ensure continued accuracy.

The following procedures are provided in the event that

recalibration becomes necessary.

For electronic calibration the only equipment required is a tweaker

or small screwdriver.

For wet calibrating a pH Monitor/controller the only items required

is a tweaker or small screwdriver, and pH buffer solutions; 4, 7

and/or 10.

pH 7 Buffer solution is recommended to wet calibrate ORP ZERO.

The electronic calibration is recommended for SPAN calibration

and adjustment, as described in “SPAN (GAIN/SLOPE)

Calibration” section V.C.1.b.

Refer to Figure V.A.1. to locate the components described in this

section.

NOTE: When opening instrument, remove front cover with care;

a ribbon cable connects the front panel and main board.

When performing calibration procedures, the

technician must take extreme care to avoid

contacting the circuitry other than the Calibration

controls. Failure to do so could result in damage to

the equipment, property and/or personal injury.

1. ELECTRONIC CALIBRATION (CIRCUIT ONLY)

This method is faster than the following method using buffer

solution, but it does not include the sensor. Therefore, it should

be used only in applications where the chance of sensor

contamination or damage are negligible.

The DISplay calibration controls have been SET and SEALED

prior to leaving the factory, and under normal conditions will not

require readjustment. This allows the customer to use the panel

meter/display to easily set the electronic calibration as described

below.

NOTE: If the DISplay calibration controls have been adjusted it

will be necessary to calibrate the RECORDER Output as

described in “0-10VDC RECORDER OUTPUT Calibration” section

V.C.2.

a. ZERO Calibration - 7pH or 0mV

1. Turn power OFF.

2. Using a standard slot screwdriver remove the

screws on the front panel.

3. Carefully wiggle the front panel to loosen the gasket and

pull gently toward you. Do not pull more than about 8

inches/20CM or you could damage the wiring harness.

4. Turn the front panel around so that the back side is

facing you and set aside.

5. Turn power ON.

6. Press and hold the ZERO Test Switch. The display

should indicate 7.00pH or 0mV (ORP). If not, set to

7.00pH or 0mV with the ZERO calibration control.

7. Turn power OFF

8. Continue with SPAN Calibration, or re-install front panel

as described in “REASSEMBLY”, page 54.

b. SPAN (GAIN/SLOPE) Calibration

1. Press and hold the SPAN Test Switch. The display

should indicate 0pH or -1999mV (ORP). If not, set to 0pH

or -1999mV with the SPAN calibration control.

2. End of calibration, turn power OFF

3. Re-install front panel as described below in

“REASSEMBLY”.

2. 0-10VDC RECORDER OUTPUT Calibration

For recorder output voltages other than 0-10VDC, substitute as

required, i.e. 5.00VDC = 2.50VDC per Section II.F.2.

pH Models

On pH models, the ZERO (7.00pH) test switch and the ZERO

calibration control are used to adjust the recorder output, 7.00pH

equals 5.00VDC. The display is then adjusted to indicate 7.00pH

with the DISplay calibration control. The SPAN test switch and

the SPAN calibration control are used to adjust the span of the

recorder output to 0VDC (0pH). Refer to figure V.A.1 for location

of items described below. A DVM set to DC Volts is required.

1. Ensure power is OFF.

2. Attach DVM to RECORDER output connection.

3. Turn power ON.

4. Press and hold the ZERO test switch.

5. The DVM should indicate 5.00VDC.

6. If not, adjust ZERO calibration control until DVM

indicates 5.00VDC.

7. Verify the panel meter/display, it should indicate 7.00pH.

8. If not, adjust DISplay control until display indicates

7.00pH.

9. Press and hold the SPAN test switch.

10. The DVM should indicate 0VDC.

11. If not, adjust SPAN calibration control until DVM

indicates 0VDC.

12. Display does not require further adjustment.

13. Turn power OFF.

14. Re-install front panel as described in

“REASSEMBLY”, page 54.

ORP Models

On ORP models, the ZERO (0mV) test switch and the ZERO

calibration control are used to adjust the recorder output, 0mV

equals 5.00VDC. The display is adjusted to indicate 0mV with the

ORP ZERO calibration control. The SPAN test switch and the

SPAN calibration control are used to adjust the span of the

recorder output to 0VDC (-1999mV). The display is then adjusted

to indicate -1999mV with the DISplay calibration control. Refer to

!

CAUTION - READ FOLLOWING CAREFULLY

Advertising
This manual is related to the following products: