Switch / o-ring assembly – Myron L 729II User Manual

Page 24

Advertising
background image

Front Panel

WARNING: there are two (2) O-rings installed on the switches,

one (1) on the shank and the other is under the push button. Both

of these O-rings must be reinstalled to maintain IP65/NEMA 4X

ratings. See figure III.A.3.

1-4 assumes this is a conversion from a single alarm/control.

Skip to #5 if new installation or assembly.

1. Carefully remove the RED LED from the front panel by

pulling lightly on the wires. It may be necessary to use a

small flat screwdriver to slightly spread the retaining

ring.

2. Remove the set point switch from the front panel by

rotating the round bezel nut CCW on the front panel, see

figure III.A.4.

3. Remove the LOWER front label.

4. Clean off remaining adhesive, if possible with alcohol.

5. Install DUAL alarm LOWER label, P# L2ALARM.

6. Install the SCO Second Alarm/Control RED LED in the

UPPER position as shown in figures III.A.5 & III.A.6.

7. Install the SCO Second Alarm/Control Set Point Switch

in the UPPER center position.

8. Re-install #1 (original) Set Point Switch in the LOWER

center position.

9. Re-install the #1 (original) RED LED in the LOWER

position.

10. The GREEN LED stays in the same location.

a. Set Point Conversion (SPC) /

Reversing Set Point

The alarm/control circuit(s) on all 720 Series II pH and ORP

Monitor/controllers are configured to trigger the alarm relay as the

reading increases.

If the user’s application requires it, the alarm/control circuit may

be easily reconfigured to trigger the alarm/control relay as the pH

(or ORP) reading decreases. Refer to figure III.A.4 for the

locations of the SPC jumpers referred to in this section.

20

Figure III.A.2

SET POINT #1

CONVERSION

FUSE*

115/

230

TRANSFORMER

RELAY #1

}

COM

{

POWER

AC LINE/ +DC
AC NEUTRAL/ -DC

GROUND

NC

NO

SOLID STATE (24VDC 30mA) OUTPUT

PA™ PIEZO ELECTRIC ALARM OR

RA™ REMOTE ALARM OR

CUSTOMER CONNECTION

PA

REMOVE JUMPER TO INSTALL SC/SCO

SECOND ALARM/CONTROL MODULE

721 726

722 727

723 728

724 729

MYRON L

COMPANY

UP

}

3S

INC

SPC

CONFIGURATION

MODULE

SPAN

ZERO

HYS1

SP1

ZERO

DO NOT

ADJUST

pH

SET POINT #1 (high)

ADJUST

SET POINT #1

HYSTERESIS

RIGHT INC

LEFT DEC

7 pH/ZERO mV

PUSH TO TEST

SPAN (GAIN )

PUSH TO TEST

DIS

ORP ZERO

PWR

C GD NC NO CM

SP

AN

DEC

GND

CHS

RA

PA

RE

MO

VE

T

O

SE

CO

ND

R

EL

AY

IN

ST

AL

L

pH

}

BK WT RD GN NU R- R+

SWITCH

FRONT PANEL

BEZEL

PUSH

BUTTON

O-RING

O-RING

Switch / O-ring Assembly