Switch and o-ring assembly – Myron L 729II User Manual

Page 39

Advertising
background image

by pulling lightly on the wires. It may be necessary

to use a small flat screwdriver to slightly spread the

retaining ring. See figure III.D.1.

b. Remove Set Point Switch from the front panel,

figure III.D.1, by rotating the round bezel nut CCW

on the front as shown in figure III.D.2.

c. Remove SCH harness from SC Module by on pulling

harness.

d. Set SCH harness aside, it is no longer used.

4. Remove Primary or Main CB alarm/control LED and Set

Point Switch from lower location on front panel.

a. Carefully remove the RED LED from the front panel

by pulling lightly on the wires. It may be necessary

to use a small flat screwdriver to slightly spread

the retaining ring. See figure III.D.1.

b. Remove Set Point Switch from the front panel, figure

III.D.1, by rotating the round bezel nut CCW on the

front as shown in figure III.D.2.

5. Re-install Primary or Main CB alarm/control LED and Set

Point Switch into UPPER location as shown in figure

III.D.6.

6. Install TH/THO harness to front panel. See figure III.D.6.

a. Install Set Point Switch into the LOWER CENTER

switch location.

b. Press RED LED into the LOWER location.

7. Remove SPC jumpers (2) from SC/SCO Second

Alarm/Control Module, see figure III.D.3 for SPC jumper

location.

8. Install TH/THO harness to SC/SCO Second

Alarm/Control Module.

a. Attach BLACK 90° connector with four wires to SPC

location with the connector 90° to the SC as shown

in figures III.D.7 & III.D.8.

b. Press WHITE connector with two wires and one

jumper into the SC - it is directional, as shown in

figures III.D.7 & III.D.8.

9. Turn SCO Set Point control fully counter clockwise - it

may be 30 turns - or until it clicks. See figure III.D.7 for

location.

10. The GREEN LED stays in the same location.

11. Locate and remove JUMPER from the Temperature

(TPO) Module as shown in figure III.D.5.

12. Attach Brown 5 wire connector to the TPO Module as

shown in figure III.D.9.

13. Add TEMPERATURE label to LOWER front panel

covering the “LOW” as shown in figure III.D.10.

14. Attach control leads to RELAY #2 as shown in figure

III.D.3.

15. Continue to Set Point Conversion or skip to “SPECIFIC

OPTION” Module Set Point adjust.

WARNING: There are two (2) O-rings installed on the switch,

one(1) on the threaded shank and the other is under the push

button. Both of these must be reinstalled to maintain IP65/NEMA

4X ratings. See figure III.D.4.

b. LOW Primary Control

If the front panel has all ready been removed from the enclosure

skip to step 3.

1. Using a standard slot screwdriver remove the

screws on the front panel.

2. Carefully wiggle the front panel to loosen the gasket and

pull gently toward you. Do not pull more than about 8

inches/20CM or you could damage the wiring harness.

3. Remove SCH harness from front panel, figure III.D.11,

and SC/SCO Second Alarm/Control Module, figure III.D.3

for locations.

a. Carefully remove the UPPER RED LED from the

front panel by pulling lightly on the wires. It may be

necessary to use a small flat screwdriver to slightly

spread the retaining ring. See figure III.D.1.

b. Remove UPPER Set Point Switch from the front

panel by rotating the round bezel nut CCW on the

35

SPC

INC

PA

SCO SET POINT #2 ADJUST

(TURN CONTROL FULL CCW)

CM

NO NC

Figure III.D.3

SC/SCO MODULE

(SECOND ALARM/CONTROL)

MAIN CIRCUIT BOARD NOT SHOWN FOR CLARITY

COM

RELAY #2

}

NO

NC

REMOVE SCH HARNESS

REMOVE BOTH SPC JUMPERS

SWITCH

FRONT PANEL

BEZEL

PUSH

BUTTON

O-RING

O-RING

SWITCH and O-RING ASSEMBLY

Figure III.D.4

BR

B

L O

R

- +

SPC

Figure III.D.5

TEMP

CONTROL

HARNESS

FS

DISPLAY

SELECT

TEMP

SENSOR

BRN

BLU

INC

INC

DEC DEC

JUMPER

REMOVE TO INSTALL

CONTROL HARNESS

TP/TPO MODULE

+

0-5VDC

-

Advertising
This manual is related to the following products: