Myron L 729II User Manual

Page 35

Advertising
background image

b. Temperature Set Point Adjust

1. Ensure power OFF.

2. Ensure the SCO Module Set Point control is set to FULL

counter clockwise position. If not, turn SCO Set Point

control fully CCW - it may be 30 turns - or until it clicks.

For location see figure III.C.13.

3. Turn power ON.

4. Press BOTH the Temperature Select switch and the Set

Point switch on the front panel.

5. Adjust the TP/TPO Module Set Point Two (SP2) to

temperature as desired. See figure III.C.1.

6. Turn power OFF.

c. Hysteresis

Hysteresis is fixed at ~±6°C unless Special Ordered with

adjustable control.

Hysteresis with adjustable control is variable from ±0.6 to 6°C,

see figure III.C.1 for location.

CAUTION: Adjusting the hysteresis too narrow may cause the

alarm to fluctuate (on-off) due to temperature variations causing

the relay to chatter. This condition is to be avoided, it may

damage the heater, chiller, etc. and will eventually damage the

relay.

REASSEMBLY

1. Carefully reinstall the front panel, bottom first. Ensure no

wires have been pinched between enclosure and front

panel.

2. Reinstall the screws and tighten.

3. To operate, turn power ON.

31

FS

INC

INC

TEMP

CONTROL

HARNESS

SET POINT

CONVERSION (SPC)

DEC DEC

SPC

Figure III.C.12

SPC

INC

PA

SCO SET POINT #2 ADJUST

(TURN CONTROL FULL CCW)

CM

NO NC

Figure III.C.13

SC/SCO MODULE

(SECOND ALARM/CONTROL)

MAIN CIRCUIT BOARD NOT SHOWN FOR CLARITY

COM

RELAY #2

}

NO

NC