Myron L 729II User Manual

Page 18

Advertising
background image

controller set point until the alarm/piezo sounds off. If

controller is not yet connected to a functioning sensor, it

may be necessary to press and hold the ZERO test

switch.

The black button on the front of the RA will mute the

piezo alarm for approximately three minutes or until you

improve the water quality (readjust controller set point).

The piezo alarm will continue to sound off every three

minutes until the user has improved the alarm condition

inside the controller. If three minutes muting is fine for

your application, skip to step 9.

8. If three minutes is too long or too short, adjust time delay

control inside RA until desired mute time is achieved

(adjustable from approximately 6 seconds to 10

minutes).

9. Replace the bottom of the RA, and secure RA to the

surface you have selected for its installation.

NOTE: If the RA does not sound off;

1. Check the polarity of the extension wire connections.

2. Be sure the controller is actually switching (relay will

click).

c. Connect to your own alarm

Use the following as guidelines.

Connector is a standard 2 wire Methode* style

connector. Connector with 8” wires, part #RAH, is

available from the Myron L Company.

Ensure your requirements do not exceed the 24 VDC

Unregulated 30mA maximum.

Ensure the polarity is correct (RED is positive), see

figure V.A.1.

Attach wires to RA.

Attach connector to controller connector (RA) per figure

V.A.1.

*Methode is registered trademark of Methode Electronics, Inc.

5. ALARM/CONTROL RELAY CONNECTIONS

Myron L Company Monitor/controllers are equipped with a “Dry

Contact” relay which is designed to energize/de-energize when

the set point is crossed. (See section IV.C.3 for set point

adjustment procedure) The relay energizes on increasing or

decreasing readings as set by the user, see section IV.C.1.

When energized (above set point), the Common (CM) will

disconnect from the Normally Closed (NC) contact and connect to

the Normally Open (NO) contact. Devices may be operated using

either the Normally Open contact or Normally Closed contact; or

both relay contacts may be used to control two devices of the

same voltage.

NOTE: A flow switch may be installed (electrically connected)

utilizing one of two convenient methods; the 3S connection, see

figure II.E.2, (remove jumper and connect flow switch across

terminals - 8” harness (RAH) available from the Myron L

Company), or inline with either relay connection, see figure II.E.3

or II.E.4.

WARNING: CONNECTING BOTH POWER SOURCE

LEADS

TO

THE

RELAY

TERMINAL

BLOCK

CONNECTORS WILL DAMAGE THE CIRCUIT BOARD

AND MAY CAUSE PERSONAL INJURY.

1. Place the user supplied Alarm relay interface cable and

watertight cable restraint into the enclosure’s

appropriate access hole. Skip for OEM.

2. Neatly connect the relay interface cable wires to the

Monitor/controller’s terminal block connectors, see

figures II.E.1, II.E.4 or II.E.5.

CAUTION: The connectors require only a small screwdriver or a

pen to push on the release levers. The release levers may be

broken or damaged if not pushed straight toward the circuit board.

DO NOT push the release levers sideways.

The easiest method of connecting the relay is shown in figures

II.E.1, and II.E.3. These show how the dry contact relay can use

incoming power to activate a controlled device (alarm, solenoid

valve, etc.) of 10 amps or less.

For 24 VAC applications, the Myron L Company offers a 115 VAC

to 24 VAC transformer, Model #VR, see figure II.E.4. Other

voltages must be user-supplied.

6. CONNECTING DISPLAY HARNESS TO DISPLAY
If the installation required the removal of the display harness from

the display (OEM installation requires connection), the following

procedure will ensure it is reinstalled without damaging the

display.

WARNING: THE DISPLAY WILL BE IRREPARABLY

DAMAGED IF THE HARNESS IS INSTALLED UPSIDE-

DOWN OR MISALIGNED. THE HARNESS MUST BE

INSTALLED AS SHOWN IN FIGURE II.E.5.

14

Figure II.E.3

DEVICE

TO BE

CONTROLLED

POWER

SOURCE

COM

NO

CM

NO

NC

RELAY

RELAY

TERMINAL BLOCK

MONITOR/CONTROLLER

CIRCUIT BOARD

OPT.

FLOW

SWITCH

X

Figure II.E.4

DEVICE

TO BE

CONTROLLED

COM

NO

115VAC

POWER

SOURCE

CM

NO

NC

RELAY

VR

RELAY

TERMINAL BLOCK

115 to 24VAC

TRANSFORMER

MONITOR/CONTROLLER

CIRCUIT BOARD

OPT.

FLOW

SWITCH

X