If you read nothing else in this

manual please read this

Quick Reference Guide.

PLEASE READ and COMPREHEND ALL

WARNINGS, CAUTIONS and ADVISEMENTS

CONTAINED WITHIN THIS MANUAL.

Failure to comply is beyond the responsibility of

the Myron L Company.

WARNING:

ALL

MONITOR/CONTROLLERS

ARE

FACTORY SET TO OPERATE ON 115 VAC. BEFORE

APPLYING POWER ENSURE THE INPUT POWER

“115/230 VAC” SELECTION IS CORRECT FOR YOUR

REQUIREMENTS. FAILURE TO DO SO IS BEYOND

THE RESPONSIBILITY OF THE MYRON L COMPANY.

See section II.E.2. and figure II.E.1.

WARNING:

ENSURE

POWER

IS

OFF

WHILE

INSTALLING

ELECTRICAL

EQUIPMENT.

IF

MONITOR/CONTROLLER

IS

INSTALLED,

ENSURE

THE POWER IS OFF BEFORE SERVICING. FAILURE

TO DO SO COULD CAUSE DAMAGE TO THE

INSTRUMENT, AND COULD BE HARMFUL OR FATAL

TO PERSONNEL. ONLY QUALIFIED PERSONNEL

SHOULD

INSTALL

OR

SERVICE

ELECTRICAL

EQUIPMENT.

WARNING: THE DISPLAY WILL BE IRREPARABLY

DAMAGED IF THE DISPLAY HARNESS IS INSTALLED

UPSIDE-DOWN OR MISALIGNED. THE HARNESS

MUST BE INSTALLED AS SHOWN IN FIGURE II.E.5.

CAUTIONS:

Before installation, ensure you have the correct

model (with options), AND it is ranged for your

application. See sections I.A., I.B. & I.G.

Do you have the correct sensor? See section I.E.

Mounting requirements. What is needed? See

section II.B.

Isolated 24 VDC output is referenced to the 0-10

VDC output. To maintain the isolation, do NOT

ground. See section II.B.4.

The following will give the installer and user a quick

overview. See the sections listed for details.

REMOVING FRONT PANEL

NOTE: When opening instrument, remove front cover with care;

a ribbon cable connects the front panel and main board.

1. Ensure power is OFF.

2. Remove the screws on the front panel.

3. Carefully wiggle the front panel to loosen the gasket and

pull gently toward you. Do not pull more than about 8

inches/20CM or you could damage the wiring harness.

REASSEMBLY

1. Carefully reinstall the front panel, bottom first.

Ensure no wires have been pinched between enclosure

and front panel.

2. Reinstall the screws and tighten.

3. To operate, turn power ON.

INTRODUCTION - Section I.

This section covers the specifications of your new

Monitor/controller including sensor information.

INSTALLATION - Section II.

This section covers how to install your new Monitor/controller;

mechanically and electrically.

OPTIONS & ACCESSORIES - Section III.

This section covers the specifications, installation, set up, and

operation of each option.

QUICK LOCATOR

SC/SCO MODULE, (Second Relay), see section III.A.

4A/4AO MODULE (4-20mA), see section III.B.

TP/TPO MODULE (Temperature), see section III.C.

TH/THO MODULE (Alarm /control Harness), see section III.D.

DUAL (stacking) Temperature (TPO) & 4-20mA (4A/4AO), see

section III.E.

PA/PAO (Piezo Alarm), see section III.F.

RA (Remote Alarm), see section III.G.

OPERATING PROCEDURES - Section IV.

This section covers a brief description of different models and

their features; how they work, and how to set them up for your

particular use.

QUICK SET POINT CONVERSION (SPC) /

REVERSING SET POINT - See Section IV.C.1.

pH & ORP Monitor/controllers are configured to trigger the alarm

relay as the reading increases.

To reverse:

1. Locate the jumper block for the alarm to be configured.

See figure V.A.1.

2. Remove and rotate the jumpers 1/4 turn and reinstall

them on their posts.

QUICK CHECK-OUT PROCEDURE -

See Section IV.C.2.

It is assumed that the Monitor/controller power is ON, that it is

connected to an appropriate Sensor, and that the Sensor is

immersed in water within the range that the Monitor/controller

will be required to read; and the front panel is removed.

1. Make a note of the reading on the display.

2. While pressing the Calibration/Full Scale Test Switch (FS

SW), verify that the front panel display is indicating a full

scale reading. If not, see Calibration, section V.C.

Continued

720 Series II MONITOR/CONTROLLER

QUICK REFERENCE GUIDE!

Advertising
This manual is related to the following products: