SENSOR INSTALLATION

The pH/ORP Sensor’s mounting orientation must provide a

continuous and adequate circulation flow to prevent the trapping

of air bubbles within the Sensor’s electrode area. Failure to do so

will result in conditions that will prevent the Sensor from

functioning properly.

1. INSERTION MODE (in-line installation)

Use one of the following as a guide. The Heavy Duty pH or ORP

sensor may be installed in any direction including inverted (sensor

UP). Substitute ORP sensor for pH where applicable.

Use approved sealant, i.e. Teflon tape as required.

1. Insert the Sensor Fitting assembly into the “T” fitting as

shown in one of the figures below.

2. Tightly secure.

2. IMMERSION OR DIP SENSOR ASSEMBLY

Use approved sealant, i.e. Teflon tape as required.

1. Insert and pull the Sensor’s cable through the extension

tube as shown in figure II.D.2.

2. Tightly attach extension tube to Sensor assembly as

shown in figure. II.D.2.

The sensor cable may be extended as necessary.

Additional cable may be ordered: Part Number pHCRD-(length)

Example:

pHCRD-300(90m)

NOTE: 1000 ft/300 meter cable lengths have been tested with no

adverse affects on the readings. However, unusual conditions

may be experienced when extended long distances, i.e. power or

ground reversal. These conditions are beyond the control and

responsibility of the Myron L Company. A clean installation is

required. If you are experiencing errors, check the

Monitor/controller with a 10 ft/3 meter cable before contacting

your distributor or the Myron L Company.

11

NOTE: HEAVY DUTY “F” MODEL SENSORS MAY BE

INSTALLED IN ANY DIRECTION INCLUDING INVERTED.

Figure II.D.2

IMMERSION OR DIP INSTALLATION

"T" FITTING

OUT

IN

CABLE

MYRON L COMPANY

Figure II.D.1.d

HEAVY DUTY "F" MODEL INSTALLED

INVERTED - BASE OF "T"

METHODS

OF

INSTALLATION

Figure II.D.1.a

M

Y

R

O

N

L

C

O

M

PA

N

Y

CABLE

"T" FITTING

INSTALLED 45° - BASE OF "T"

Figure II.D.1.c

HEAVY DUTY "F" MODEL INSTALLED

HORIZONTALLY - BASE OF "T"

CABLE

"T" FITTING

MYRON L COMPANY

MYRON L COMPANY

"T" FITTING

OUT

IN

INSTALLED VERTICALLY - BASE OF "T"

Figure II.D.1.b

CABLE

CABLE

MYRON L COMPANY

MYRON L COMPANY

COUPLING

MYRON L COMPANY

M

Y

R

O

N

L

C

O

M

P

A

N

Y

M

Y

R

O

N

L

C

O

M

P

A

N

Y

M

Y

R

O

N

L

C

O

M

P

A

N

Y

Advertising
This manual is related to the following products: