Myron L 729II User Manual

Page 4

Advertising
background image

3. Press and hold the “SET POINT” switch on the front

panel. Using a tweaker or a small screwdriver, adjust the

Set Point trimmer adjustment screw on the circuit board

to sweep the display from zero to full scale. (A digital

display may be blank at the full scale end. This is

normal.) Listen for the alarm relay to click on and off as

the alarm set point moves past the water reading.

4. Adjust the alarm to the desired set point value. Release

the “SET POINT” switch.

NOTE: For Models with SC/SCO module, repeat STEPS 3 & 4 to

check out Set Point #2.

QUICK SET POINT ADJUSTMENT -

See Section IV.C.3.

The set point setting is based upon the user’s particular water

purity specifications or requirements. NOTE: The optional

second relay/alarm is “stacked” on the first relay/alarm,

therefore, when setting the optional second relay/alarm Set

Point, the #1 Set Point must be ‘set’ first.

1. While pressing the “SET POINT” switch, turn the Set

Point #1 adjustment screw (see figure V.A.1) until the

desired set point value is indicated on the display.

HYSTERESIS (DEAD BAND) ADJUSTMENT -

See Section IV.C.4.

PRIMARY COMPONENT IDENTIFICATION -

Section V.A.

Review the figure below to familiarize yourself with the Main

circuit board assembly. The diagram has the second

alarm/control module option installed.

QUICK CALIBRATION - Section V.C.
WARNING: When performing calibration procedures,

the technician must take extreme care to avoid

contacting the circuitry other than the CALibration

control. Failure to do so could result in damage to

the equipment, property and/or personal injury.
The following assumes the front panel has been

removed and the power is ON.

ELECTRONIC CALIBRATION (CIRCUIT ONLY) -

See Section V.C.1.

ZERO Adjustment

1. Press and hold the ZERO Test switch. The display should

indicate 7 pH or 0 mV selected. If not, set to 7 pH or 0 mV

with the ZERO CALibration control.

2. Turn power OFF.

3. Re-install front panel as described in “REASSEMBLY”.

4. To operate, turn power ON.

0-10VDC Recorder Calibration - See Section V.C.2.

Using Standard Solutions - Section V.C.3.

The BEST method of verifying and recalibrating your pH or ORP

Monitor/controller is with NIST traceable Standard Solution

(available from the Myron L Company). Because it includes the

sensor, the entire system is recalibrated.

NOTE: Another means of verification or calibration of pH or ORP

models is to use the transfer standard method, using a hand-held

or portable instrument capable of these measurements, i.e. the

Myron L Ultrameter™. See section V.C.4 for description.

The following procedure describes the easiest method for

standard solution calibration of your Monitor/controller.

1. Using 7pH buffer solution rinse a clean glass beaker

thoroughly with the buffer solution.

2. Place sensor in the beaker of buffer solution. Level of

buffer solution should be high enough to cover at least 1”

above the sensor lower end.

3. Carefully shake the sensor to remove air bubbles from

inside the sensor bore hole.

4. Allow 5-10 minutes for temperature to equilibrate. For the

quickest and the best results, both the sensor and

solution should be at the same temperature.

5. Read the panel meter/display. The display should match

the value and units of measure located on the bottle of

buffer solution. If the reading is different, adjust ZERO

calibration control on the main control circuit board

until the reading is 7pH or 0mV.

6. Repeat steps 2 - 5 using either 4 or 10 pH buffer

solution.

7. If reading is incorrect, adjust SPAN calibration

control on the main control circuit board until reading

matches buffer solution.

8. After adjustment, turn power OFF.

9. Re-install front panel as described in “REASSEMBLY”.

10. To operate, turn power ON.

SENSOR SUBSTITUTE CALIBRATION -

See Section V.C.4.

TRANSFER STANDARD METHOD - See Section V.C.5.

21 may 03

{

POWER

AC LINE/ +DC
AC NEUTRAL/ -DC

GROUND

Figure V.A.1

SPAN

ZERO

DO NOT

ADJUST

RELAY #1

}

COM

NC

NO

3SO/3SE

SWITCH

CONNECTION

DISPLAY/METER

CONNECTOR

4-20 CONNECTOR

SET POINT #1

CONVERSION

TRANSFORMER

FUSE*

115/

230

ZERO

CALIBRATION

CONTROL

SET POINT #2 CONTROL

SC/SCO

OPTIONAL

SECOND

ALARM/

CONTROL

MODULE

SET POINT #2

CONVERSION

#2 RELAY

LIGHTS &

SWITCH

CONNECTOR

DISPLAY CALIBRATION

CONTROL (FACTORY SET)

RUBBER TAPE

(DO NOT

REMOVE)

SPC

INC

PA

SET POINT

#1 CONTROL

INC

SET POINT #1

HYSTERESIS

RIGHT INC

LEFT DEC

CHASSIS GROUND - OEM

INC

SET POINT #2 HYSTERESIS

RIGHT INC / LEFT DEC

3S

SPC

0-10VDC OUTPUT

B

L

K

W

H

T

R

E

D

G

R

N

N

E

U

(+)

SENSOR

(-)

+

COM

RELAY #2

}

NO

NC

+

SPAN (GAIN)

CALIBRATION

CONTROL

721 726

722 727

723 728

724 729

MYRON L
COMPANY

7 pH/ZERO mV

PUSH TO TEST

SPAN (GAIN/SLOPE)

PUSH TO TEST

CONFIGURATION

MODULE

CONFIGURATION

MODULE

ALIGNMENT

ORP ZERO CALIBRATION

CONTROL (FACTORY SET)

SOLID STATE (24VDC 30mW)

OUTPUT

PA™ PIEZO ELECTRIC ALARM

OR

RA™ REMOTE ALARM OR

CUSTOMER CONNECTION

SOLID STATE (24VDC 30mW)

OUTPUT

PA™ PIEZO ELECTRIC ALARM

OR

RA™ REMOTE ALARM OR

CUSTOMER CONNECTION

HYS1

SP1

ZERO

DIS

ORP ZERO

UP

BK WT RD GN NU R- R+

PWR

C GD NC NO CM

FLOW SWITCH

RA

PA

CM

N

O

N

C

DE

C

pH

CH

S

SP

AN

GN

D

REMOVE T

O INST

ALL

SECOND RELA

Y

}

Advertising
This manual is related to the following products: