Specifications:

As required to meet USP 25.

Installation

Briefly -

The BLACK and WHITE sensor leads are REVERSED in the

sensor connector.

Requirements:

None

WARNING: BEFORE STARTING, IF MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER

IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE

TO THE INSTRUMENT, AND COULD BE HARMFUL OR

FATAL

TO

PERSONNEL.

ONLY

QUALIFIED

PERSONNEL

SHOULD

INSTALL

OR

SERVICE

ELECTRICAL EQUIPMENT.

NOTE: When opening instrument, remove front cover with care;

a ribbon cable connects the front panel and main board. If the

front panel has all ready been removed from the enclosure skip to

#4.

1. Using a standard slot screwdriver remove the

screws on the front panel.

2. Carefully wiggle the front panel to loosen the gasket and

pull gently toward you. Do not pull more than about 8

inches/20CM or you could damage the wiring harness.

3. Turn the front panel around so that the back side is

facing you and set aside.

4. If sensor is installed;

locate and REVERSE the BLACK and the WHITE sensor

leads on the MAIN Circuit Board, as shown in figure

II.E.2.

5. If sensor is NOT installed;

install the BLACK sensor lead into the connector labeled

WT (WHITE) AND the WHITE lead into the connector

labeled BK (BLACK).

Install the remaining sensor leads as color coded.

6. To return to “normal” temperature compensation,

reinstall the sensor leads as labeled - WHITE to (WT) and

BLACK to (BK).

7. Carefully reinstall the front panel, bottom first. Ensure no

wires have been pinched between enclosure and front

panel.

8. Reinstall the screws and tighten.

9. To operate, turn power ON.

NOTE: Recalibration will require both the solution and sensor be

at 25°C for maximum accuracy.

4. ISOLATED SOLID STATE OUTPUT

CAUTION: Isolated 24 VDC output is referenced to

the 0-10 VDC output. To maintain the isolation, do

NOT ground.

24 VDC Unregulated 30mA max. The following instructions are

assuming the Monitor/controller enclosure is already open.

a. Piezo Electric Alarm - PA/PAO (option)

For additional information, see Piezo Alarm under Options in

section III.G.

1. If not already installed, peel tape backing from PIEZO

and press into place per figure III.I.3.

2. Attach connector to main circuit board per figure III.I.4.

NOTE: If remotely mounted; cut wires and splice as necessary,

use comparable wire. Piezo requires 1/4” (6.35mm) hole in user

panel.

b. Remote Alarm - RA™ (option)

For additional information, see RA Instructions under options

in section III.H.

1. Run user supplied #22, 2 conductor speaker type wire

from Monitor/controller to RA location as necessary.

Additional wire may be ordered part #RAW-200, see

Options & Accessories.

2. Open the RA by removing the four screws.

3. Locate and remove the 8” 2 conductor wire attached to

RA.

4. At the controller, connect the extension wires to the 8”

2 conductor wire with the wire nuts provided — Black to

Positive (+) and White to Negative (-). Be sure to first

pass the wire through the user supplied waterproof strain

relief in the enclosure.

5. Plug the reddish brown female connector into the male

connector on the controller CB marked either RA or PA

(see inside case label for location). It will only go on the

connector one way.

6. At the RA, connect the wires to the connector — Black

to Positive (+) and White to Negative (-).

7. To test, simply turn on the controller and adjust

13

!

CAUTION - READ FOLLOWING CAREFULLY

HYS1

SP1

SPAN

ZERO

DIS

ORP ZERO

SPAN

ZERO

Figure II.E.2

3S

pH/mV ZERO

PUSH TO TEST

INC

DEC

SPC

pH Main CB Assembly

0-10VDC

OUTPUT

S

ENSOR

LEADS

REVERSE BLACK & WHITE LEADS

SPAN (GAIN)

PUSH TO TEST

DO NOT

ADJUST

WHITE

BLACK

UP

}

ORP

}

BK WT RD GN NU R- R+

SP

AN

DE

C