Myron L 729II User Manual

Page 49

Advertising
background image

45

B.

OEM FRONT PANEL INDICATORS &

CONTROLS

The simulated OEM front panel illustrations to the right, and

switch and indicator operational descriptions have been provided

to assist the OEM user in identifying and operating the 720 Series

II

Monitor/controllers.

Refer to Section V.C. for Setup procedures, and Section V.D. for

Check-Out procedures.

1. RED “HIGH SET POINT” LED INDICATOR

The red LED indicator light is ON only when the water’s pH or ORP

reading is HIGH or ABOVE the set point adjustment. May be

reversed if desired.

2. GREEN “LOW SET POINT” LED INDICATOR

The green LED indicator light is ON only when the water’s pH or

ORP reading is LOW or BELOW the set point adjustment. May be

reversed if desired.

3. “SET POINT” SWITCH(ES)

When the “SET POINT” switch is depressed, the internal set point

reading is immediately displayed on the front panel display. See

figures IV.B.1 thru 4.

4. ANALOG METER OR DIGITAL DISPLAY

Panel mounted analog meter or digital display provide a

continuous readout of the water being monitored.

pH Models 721II & 722II and ORP Models 726II & 727II are

equipped with analog meter only as shown in figures IV.B.4 & 5.

pH Models 723II & 724II and ORP Models 728II & 729II are

equipped with a 3 1/2 digit, 1/2” Liquid Crystal Digital Display, as

shown in figures IV.B.1 thru 3., with a 3 1/2 digit backlit LCD as an

option.

5. OPTIONAL PANEL MOUNTED ITEMS

TP/TPO Module Switch

A digital Monitor/controller with the optional TP/TPO Temperature

Module has an additional switch on the front panel as shown in

figure IV.B.1.. This push-button momentary switch when pushed

gives the user a direct reading of the temperature of the solution

from 0-200°C. Additional option TH/THO includes a control

function capability. See figure IV.B.1.

7pH or ZEROmV Test Switch (TEST)

This optional feature allows the user to see the ZERO TEST value

without opening the panel to push the internal switch. See figures

IV.B.1 and IV.B.3 for typical examples.

Piezo Alarm

Audible alarm sounds off automatically when the set point is

reached. Mounted by OEM in any convenient location as shown in

figure IV.B.3.

WARNING: THE DISPLAY WILL BE IRREPARABLY

DAMAGED IF THE HARNESS IS INSTALLED UPSIDE-

DOWN. THE HARNESS MUST BE INSTALLED AS

SHOWN IN FIGURE II.E.3.

HIGH

LOW

SET POINT

HIGH

LOW

SET POINT

LOW

SET POINT

TEMPERATURE

TEMPERATURE

TEST

Figure IV.B.1

Figure IV.B.2

Figure IV.B.3

Figure IV.B.4

Figure IV.B.5

HIGH

LOW

SET POINT

TEST

NOTE: Boxes around displays

to simulate OEM panels.

0

14

12

6

2

8 10

4

pH

ORP

2

2

1.5

.5

1.5

.5 1

1

0

X1000

Analog pH

Monitor/controller

(Single Alarm/control)

Digital pH

Monitor/controller

with optional Dual

Alarm/controls

Digital ORP

Monitor/controller with

optional Front Panel

Test Switch, Single

Alarm/control, and

Piezo Alarm.

Analog

ORP

Monitor

Digital pH

Monitor/controller

with optional

Dual Alarm/controls,

and Temperature with

control function, and

Front Panel Test Switch.

726II

722II

728II-PTS-PA

723II-SC