a. TPC “Calibration” Module Procedure

This procedure assumes the front panel is removed.

1. Ensure power is OFF.

2. Remove the sensor leads from the Temperature Module.

3. Remove the 0-5VDC leads from the Temperature Module.

4. Press TP Calibration Module (TPC) firmly into sensor

input connectors as shown in figure III.C.5.

5. Attach the DVM to the 0-5VDC output on the TPC

module. See figure III.C.5.

6. Turn power ON. The DVM should indicate 0.0 volts.

7. If not, adjust the calibration control marked “ZERO” (see

figure III.C.1 for location) until the DVM indicates 0.0

volts as shown in figure III.C.5.

8. Press switch on the TPC. See figure III.C.6.

9. The DVM should indicate 5.0 volts. See figure III.C.6.

10. If not, while pressing the TPC Module switch, adjust the

calibration control marked “FS” (see figure III.C.1 for

location) until the DVM indicates 5.0 volts as shown in

figure III.C.6.

11. Press the front panel TEMPERATURE select switch and

the TPC Module switch, the display should indicate

199.9 as shown in figure III.C.7.

12. If not adjust the calibration control marked “DIS” (see

figure III.C.1 for location) until the reading is 199.9 as

shown in figure III.C.7.

13. Calibration is complete.

14. Turn power OFF.

15. Remove TPC Module by pressing on each Phoenix

connector release arm*.

16. Reconnect sensor leads as labeled.

17. Reconnect 0-5VDC output leads as labeled.

18. Continue or reinstall the front panel and tightly secure

retaining screws, see REASSEMBLY, page 31.

CAUTION: The sensor input and 0-10VDC output connectors

require only a small screwdriver or a pen to push on the release

levers. The release levers may be broken or damaged if not

pushed straight toward the CB. DO NOT push the release levers

sideways.

b. Precision Resistor Calibration Procedure

This procedure assumes the front panel is removed.

1. Ensure power is OFF.

2. Remove the sensor leads from the Temperature Module.

3. Install ZERO resistor (1000.0Ω) across sensor terminals

BL and OR. See figure III.C.8.

4. Install JUMPER between BR & BL as shown in figures

III.C.8 & III.C.9.

5. Attach the DVM to the 0-5VDC output. See figures III.C.8

& III.C.9.

6. Turn power ON. The DVM should indicate 0.0 volts.

7. If not, adjust the calibration control marked “ZERO” (see

figure III.C.1 for location).

8. Install SPAN resistor (1758.56Ω) across sensor

terminals BL and OR. See figure III.C.9.

9. The DVM should indicate 5.0 volts.

10. If not, adjust the calibration control marked “FS” (see

figure III.C.1 for location) until the DVM indicates 5.0

volts as shown in figure III.C.9.

11. Press the front panel TEMPERATURE select switch, the

display should indicate 199.9 as shown in figure

III.C.10.

12. If not adjust the calibration control marked “DIS” (see

figure III.C.1 for location) until the reading is 199.9.

13. Calibration is complete.

14. Turn power OFF.

15. Remove DVM from 0-5VDC.

16. Remove resistor and jumper by pressing on each

Phoenix connector release arm*.

17. Reconnect sensor leads as labeled.

18. Reconnect 0-5VDC output leads as labeled.

29

Figure III.C.5

DV

-
+

0.0

Press TPC Module into

TPO connector as shown.

Press TPC Module into

TPO connector as shown.

Connect DVM leads as shown.

BR BL OR - +

TPC Module

PRESS BUTTON FOR

199.9ºC/5.0VDC

MYRON L

BLUE

BROWN

OUTPUT

-

0-5VDC

+

ORANGE

BROWN BLUE

COMPANY

Figure III.C.6

DV

-
+

5.0

Press TPC Module into

TPO connector as shown.

Press TPC Module into

TPO connector as shown.

Connect DVM leads as shown.

BR BL OR - +

TPC Module

PRESS BUTTON FOR

199.9ºC/5.0VDC

MYRON L

BLUE

BROWN

OUTPUT

-

0-5VDC

+

ORANGE

BROWN BLUE

COMPANY

Figure III.C.7

199.9

FRONT PANEL DISPLAY

Advertising
This manual is related to the following products: