background image

8. Reverse the above procedure using the new scale.

NOTE: While replacing the clear plastic cover, ensure the

ZA screw pin on the cover aligns with the inverted “Y” shaped

slot on the meter movement. See figure II.G.5.

IF, the inverted “Y” shaped slot is inadvertently moved -

BEFORE installing cover:

a. Recenter the inverted “Y” to the meter case as

shown in figure II.G.5.

b. Center the Zero Adjust (ZA) PIN on the clear

plastic cover as shown in figure II.G.6.

9. Using a standard slot screwdriver, ensure the ZA is

operating - pointer swings left and right when turning the

ZA with a standard slot screwdriver. See figure II.G.7.

10. Set Meter ZERO as shown in figure II.G.7.

11. Reinstall meter to front panel, reversing steps 1-3

(except OEM models).

REASSEMBLY

1. Carefully reinstall the front panel, bottom first, ensure no

wires have been pinched.

2. Reinstall the screws and tighten.

3. To operate, turn power ON.

18

Figure II.G.6

CLEAR PLASTIC COVER

(Inside View)

CENTER Zero Adjust (ZA) PIN

Figure II.G.7

Zero Adjust (ZA)

CLEAR PLASTIC COVER

(case not shown for clarity)

0

Align Pointer with FIRST Line

Figure II.G.8

UP

PIN #10

ANALOG

REMOVED

Bottom View

ORP

CONFIGURATION

MODULE

UP

Top View

}

ORPD O

ORP

A O

“ANALOG”

Figure II.G.9

UP

PIN #12

DIGITAL

REMOVED

Bottom View

ORP "DIGITAL"

CONFIGURATION

MODULE

UP

Top View

ORP

A O

ORPD O

}