9.5.5.

COSPEED - external

This parameter affects the temperature pattern only in case of external
control.

Possible parameters: 0.0 ... 5.0

During selftuning, the control parameters Xp, Tn and Tv of a controlled
system are automatically determined and stored. Depending on the
controlled system, time for tuning can be unequally longer. This
controller layout allows protection of sensitive objects requiring
temperature application.

°C

t

Setpoint

5

3

0

Int. temperature

Ext. temperature

As soon as a co-speed factor is set, it is
considered for calculating the control
parameters. As shown in the diagram,
tuning times become shorter the higher
the co-speed factor is, but overshooting
can happen in the internal system.







9.5.6.

Control parameters – XPU-, XP-, TN-, TV- EXTERNAL

Setting range: 0.1 ...99.9

Setting range: 3 ...9999

Setting range: 0 ... 999

In most cases the control parameters preset in the factors are adequate
for achieving an optimum temperature sequence.
The control parameters allow adjustment to special control processes.

Setting range: 0.1 ... 99.9

Proportional range >Xpu<

The proportional range Xpu of the cascaded controller is only needed for
external control.