9.5.2.

SELFTUNING

Selftuning:

When performing a selftuning for the controlled system (temperature
application system), the control parameters Xp, Tn and Tv are
automatically determined and stored.

Possible parameters:

OFF - no selftuning

The control parameters ascertained during the last identification are

used for control purposes.

ONCE - single selftuning (factory setting)

The instrument performs a single selftuning of the controlled system

after each start with the

key or after receiving a start

command via the interface.

ALWAYS - continual selftuning

The instrument performs a selftuning of the controlled system

whenever a new setpoint is to be reached.

Use this setting only when the temperature application system

changes permanently.


9.5.3.

Dynamic internal

°C

t

APER

NORM

Setpoint

temp. stability

°C

t

APER

NORM

temperature ramp

temp. stability

This parameter affects the temperature sequence only in
case of internal control.

Factory setting:

APER (aperiodic)

Possible parameters:

NORM Allows for reaching the setpoint faster – with

setpoint change or ramp function – but
overshooting of up to 5 % is possible.

APER

Ramp function: the increase of temperature
occurs temporally offset and achieves the
target temperature without overshooting.
Setpoint change: The temperature increases at
the same rate, the target temperature is
achieved without overshooting.

With both settings constant temperature is achieved
after approximately the same time.