Switch-over from warning to shutdown function – JULABO SE-Z Heating Circulators User Manual

Page 29

Advertising
background image

HE / SE

29

8.2.

Switch-over from warning to shutdown function

If a shutdown of functional elements (e.g. heater, circulating pump) is
required when the limit values are exceeded or undercut the circulator can
be changed over from warning function >WARNING< to shutdown function
>ALARM<.

Factory setting:
>WARNING<

1. Press the key

.

2. Select the menu >SECVAL -AL-TYPE< by pressing the

key.

3. Press the

key and the set parameter will flash

.

(Example: WARNING)

4. Change the parameter by pressing the

key and confirm by

pressing the

key.

or

press the

key if the parameter is to retained.


Setting >WARNING<
A mere warning function with optical and audible warning signal (interval
tone) A message appears on the VFD COMFORT-DISPLAY:

or

OVERTMP

SUBTEMP

• Setting >ALARM<

Temperature limit with shutdown of heater and circulating pump.
An audible alarm sounds (continuous tone) and a message appears on
the VFD COMFORT-DISPLAY:

or

OVERTMP

SUBTEMP