Alarm-output / stand-by-input – JULABO SE-Z Heating Circulators User Manual

Page 64

Advertising
background image

Menu functions

64

Example out of diagram A:

• Adjusting the voltage source for an output of 7.6 V!

The instrument calculates this value from the rise angle of the two
predecided end points (in example A: 7.6 V correspond to an external
setpoint temperature of 152.0

°C ).

After returning the VFD display to standard display, this value is
displayed in line 2 (Example: SP 152.00

°C).


Notice:

If this adjustment is not correctly performed at two different points, the setpoint setting
will be incorrect.

Important:
The usable temperature range between > LVAL < and > UVAL < is limited to the
configured working temperature range of the circulator resp. of the unit combination
(working temperature range see technical specifications)

9.9.3.

ALARM-output / Stand-by-input

ALARM

Alarm output (10)

(for external alarm signal)

Possible parameters:

(STANDBY or ALARM or AL-STBY)

This socket is a potential-free change-over contact. With the adjustments
in the menu item > AL-OUT Function< all operating conditions can be
signaled without having to change the pin assignments.

Signification of the terms under menu item >Function< :

The circulator is in condition >STANDBY< or >ALARM< or >AL-STBY< -
both conditions are signaled.

Alarm-TYPE

Possible parameters:

( NORMAL or INVERS )

For >TYPE- NORMAL< pins 2 and 3 are connected in any case
according to the selected > AL-OUT Function<

For >TYPE- INVERS < pins 2 and 1 are connected in any case according
to the selected > AL-OUT Function<

Switching capacity

max. 30 W / 40 VA

Switching voltage

max. 125 V

∼/−

Switching current

max. 1 A

Advertising