3.3.

Safety recommendations

Follow the safety instructions to avoid personal injury and property damage. Also, the
valid safety instructions for workplaces must be followed.

• Only connect the unit to a power socket with an earthing contact (PE – protective

earth)!

• The power supply plug serves as a safe disconnecting device from the line and

must always be easily accessible.

• Place the unit on an even surface on a base made of nonflammable material.

• Do not stay in the area below the unit.

• Make sure you read and understand all instructions and safety precautions listed in

this manual before installing or operating your unit.

• Set the excess temperature safety installation at least 25 °C below the fire point of

the bath fluid.

• Observe the limited working temperature range when using plastic bath tanks.

• Never operate the unit without bath fluid in the bath.

• Pay attention to the thermal expansion of bath oil during heating to avoid

overflowing of the fluid.

• Prevent water from entering the hot bath oil.

• Do not drain the bath fluid while it is hot!

Check the temperature of the bath fluid prior to draining (e.g., by switching the unit
on for a short moment).

• Use suitable connecting tubing.

• Avoid sharp bends in the tubing, and maintain a sufficient distance from

surrounding walls.

• Make sure that the tubing is securely attached.

• Regularly check the tubing for material defects (e.g., for cracks).

• Never operate damaged or leaking units.

• Always turn off the unit and disconnect the mains cable from the power source

before performing any service or maintenance procedures, or before moving the
unit.

• Always turn off the unit and disconnect the mains cable from the power source

before cleaning the unit.

• Always empty the bath before moving the unit.

• Transport the unit with care.

• Sudden jolts or drops may cause damage in the interior of the unit.

• Observe all warning labels.

• Never remove warning labels.

• Never operate units with damaged mains power cables.

• Repairs are to be carried out only by qualified service personnel.

 Some parts of the bath tank and the pump connections may become extremely hot

during continuous operation. Therefore, exercise particular caution when touching
these parts.