Installation of electronic module

76

Error messages

Description

-23 WARNING: HIGH TEMPERATURE

ON COMPRESSOR STAGE 1

Excess temperature on compressor stage 1.

-24 WARNING: HIGH TEMPERATURE

ON COMPRESSOR STAGE 2

Excess temperature on compressor stage 2.

-25 REFRIGERATOR WARNING

Error in the cooling machine.

-26 WARNING: STAND-BY PLUG IS

MISSING

External standby contact is open.
(see page 64 and 70)

-30 CONFIGURATION ERROR:

CONFIRM BY PRESSING <OK> ON
CIRCULATOR

The configuration of the circulator does not conform to its
present use.
Press >OK< to automatically perform a single modification
of the configuration.

-33 SAFETY SENSOR ALARM

Excess temperature sensor short-circuited or interrupted.

-38 EXTERNAL SENSOR SETPOINT

PROGRAMMING ALARM

Ext. Pt100 sensor input without signal and setpoint
programming set to external Pt100.

-40 NIVEAU LEVEL WARNUNG

Low liquid level warning in the internal reservoir.













13. Installation of electronic module

Caution:

Always turn off the unit and disconnect the mains cable from the power source before
performing any service or maintenance procedures, or before moving the unit.
Configuration, installation, maintenance and repairs on the circulator may only be
carried out by qualified personnel.

ALARM

REG+
E-PROG

STAND-BY

A

B

• Turn off the unit with the mains switch and disconnect the

power plug.

• Open the screws (A) and remove the dummy plate.
• Carefully insert the upper edge (B) into the rear panel from

below, align the electronic module at top and bottom and
applying light pressure connect the 15-pole connector.

• Attach the electronic module using the screws (A).
• The circulator is ready for use. It automatically recognizes the

installed electronic module.