Menu functions

44

9.5.1.

CONTROL – Control INTERNAL / EXTERNAL

Pt100

Switchover can only be effected if a Pt100 external sensor is
connected.

Factory setting: INT

IMPORTANT: Additional measures for external temperature control

Suggested settings for external temperature control:
BAND HIGH / LOW and INTERN MAX / MIN
see chapter >MENU LIMITS<

Sensor calibration of the Pt100 external sensor is carried out in the
menu >ADJUST<, submenu >ATC SENOR - EXT<; set ATC
STATUS< to >OFF<
(See page 49).

Attention:

Place the external sensor into the temperature-controlled medium and securely fix the
sensor.

Pt100

M+R

Accessory: Pt100 external sensor

Order No.

Description

Material

Cable

8981003

200x6 mm Ø,

stainless steel

1.5 m

8981005

200x6 mm Ø,

glass

1.5 m

8981006

20x2 mm Ø,

stainless steel

1.5 m

8981010

300x6 mm Ø,

stainless steel

1.5 m

8981015

300x6 mm Ø,

stainless steel / PTFE coated

3 m

8981013

600x6 mm Ø,

stainless steel / PTFE coated

3 m

8981016

900x6 mm Ø,

stainless steel / PTFE coated

3 m

8981014

1200x6 mm Ø,

stainless steel / PTFE coated

3 m

8981103 Extension cable for Pt100 sensor

3.5 m

8981020 M+R in-line Pt100 sensor

The M+R in-line Pt100 sensor is a flow sensor and can be installed loop
circuit