Input of the connector - reg+e-prog – JULABO SE-Z Heating Circulators User Manual

Page 62

Advertising
background image

Menu functions

62

9.9.2.

Input of the connector - REG+E-PROG

CHANNEL 1

CHANNEL 3

CHANNEL 2

REG+E-PROG

EPROG

Setting needs to be carried out, if

1.

setpoint programming is to be made via an external voltage or
current source or programmer.
For this, in the menu > MENU / CONFIG < first set
the menu item > SETP < to >EPROG<.

>

>

2.

the heater variable should be controlled via an external control pulse.
For this, in the menu > MENU / CONFIG < set
the menu item >ACTVAR< to >EPROG<.

>

>

(refer to page 41).

3.

the signal of an external flow meter should be registered.
>FLOWRAT< see below.

>FLOWRAT<
Indication on display

FLOWRAT
The E-Prog input can only be used either under menu item >CONFIG
-SP EXT< < or under menu item >CONFIG ACTVAR<
If the input is neither occupied by > SP EXT < or > ACTVAR <, the
signal of a flow meter can be connected.

First set >MENU CONFIG<

>

>

then >MENU IN/OUT<

>

>





Beispiel:

Settings on input - E-PROG

1.

Connect the external voltage or current source or programmer to
socket REG+E-PROG of the circulator.

2.

Selecting the signal:
The programmer (E-PROG) input of the circulator can be matched to
the output signal of the external voltage or current source.

voltage input = U

current input = I

3.

Define the input variable:

Setpoint programmed by external voltage or current

Manipulated (ACTuated) variable for the heater

with an external control pulse

Signal of an external flow sensor