BAND HIGH / LOW – Band limitation

The band limitation is active during external control. Varied, practice-
oriented setting are feasible for heat-up and cool-down phases.

Setting range: 0 °C ... 200 °C

BAND HIGH and BAND LOW allow for the limitation of the difference
between the temperatures in the internal bath and the external system to
any maximum value for the heat-up and cool-down phase.
During the heat-up phase this difference value is always added to the
actual external temperature. During the cool-down phase the difference
value is subtracted.

25

50

10 20 30 40 50 60 10 20 30 40 50 60 10 [min]

75

100

150

200

°C

t

external system

Internal bath

BAND HIGH

BAND LOW

Sense of a band limitation:

Protection of objects and samples
by gentle temperature control

Protection of e.g. glass reactors
from thermal shock.