Bath fluids – JULABO SE-Z Heating Circulators User Manual

Page 19

Advertising
background image

HE / SE

19

5.2.

Bath fluids

Caution:

Carefully read the safety data sheet of the bath fluid used, particularly with regard to
the fire point!
If a bath fluid with a fire point of

≤65 °C is used, only supervised operation is possible.

Water: The quality of water depends on local conditions.
• Due to the high concentration of lime, hard water is not suitable for temperature

control because it leads to calcification in the bath.

• Ferrous water can cause corrosion - even on stainless steel.
• Chloric water can cause pitting corrosion.
• Distilled and deionized water is unsuitable. Their special properties cause

corrosion in the bath, even in stainless steel.

Recommended bath fluids:

Bath fluid

Temperature range

soft/decalcified water

5 °C to 80 °C

JULABO bath fluids

JULABO
Description

Thermal
G

Thermal
M

Thermal
HS

Order Number

10 liters

8 940 124

8 940 100

8 940 102

5 liters

8 940 125

8 940 101

8 940 103

Temperature range

°C

-30 ... 80

40 ... 170

20 ...250

Flash point

°C

--

284

270

Fire point

°C

--

306

>360

Color

light yellow

clear

light brown

JULABO
Description

Thermal
H10

Thermal
H20S

Thermal
H200

Order Number

10 liters

8 940 114

8 940 108

8 940 134

5 liters

8 940 115

8 940 109

8 940 135

Temperature range

°C

-20 ... 180

0 ... 220

60 ...200

Flash point

°C

190

230

292

Fire point

°C

216

274

334

Color

clear

light brown

clear

See website for list of recommended bath fluids.

ATTENTION:

The maximum permissible viscosity is 70 mm

2

/s.

Caution:

Fire or other dangers when using bath fluids that are not recommended:
Please contact JULABO before using other than recommended bath fluids.
Use only nonacidic and noncorrosive bath fluids.

JULABO assumes no liability for damage caused by the selection of an unsuitable
bath liquid.
Unsuitable bath fluids are fluids which, e.g.,
• are highly viscous

(much higher than recommended at the respective working temperature)

Advertising