12.3. List of commands

out commands: Setting temperature values or parameters.

Command

Parameter

Response of circulator

out_mode_01

0

Use working temperature > SETPNT 1<

out_mode_01

1

Use working temperature > SETPNT 2<

out_mode_01

2

Use working temperature > SETPNT 3<

out_mode_02

0

Selftuning „off“.
Temperature control by using the stored parameters.

out_mode_02

1

Selftuning „once“
Single selftuning of controlled system after the next start.

out_mode_02

2

Selftuning „always“
Continual selftuning of controlled system whenever a new
setpoint is to be reached.

out_mode_03

0

Set external programmer input to voltage.
Voltage

0 V ... 10 V

out_mode_03

1

Set external programmer input to current.
Current

0 mA ... 20 mA

out_mode_04

0

Temperature control of internal bath.

out_mode_04

1

External control with Pt100 sensor.

out_mode_05

0

Stop the unit = R –OFF-.

out_mode_05

1

Start the unit.

out_mode_08

0

Set the control dynamics - aperiodic

out_mode_08

1

Set the control dynamics - standard

out_sp_00

xxx.xx

Set working temperature. „SETPNT 1“

out_sp_01

xxx.xx

Set working temperature. „SETPNT 2“

out_sp_02

xxx.xx

Set working temperature. „SETPNT 3“

out_sp_03

xxx.xx

Set high temperature warning limit „OverTemp“

out_sp_04

xxx.xx

Set low temperature warning limit

„SubTemp“

out_sp_06

xxx.xx

Set manipulated variable for the heater via serial interface
-99.99 ... +100 [%]

out_sp_07

x

Set the pump pressure stage. (1 ... 4)

out_par_04

x.x

CoSpeed for external control 0 ... 5.0

out_par_06

xxx

Xp control parameter of the internal controller.

0.1 ... 99.9

out_par_07

xxx

Tn control parameter of the internal controller.

3 ... 9999

out_par_08

xxx

Tv control parameter of the internal controller.

0 ... 999

out_par_09

xxx

Xp control parameter of the cascade controller.

0.1 ... 99.9

out_par_10

xxx

Proportional portion of the cascade controller.

1 ... 99.9

out_par_11

xxx

Tn control parameter of the cascade controller.

3 ... 9999