Safety installations, warning functions

28

8.1.1.

Early warning system, low level protection

This low level protection is independent of the control circuit and is
divided into two sections:

1. Switch in stage 1 recognizes a defined fluid level

.

An audible warning sounds (interval tone) and together with the
ticker: > LOW LEVEL WARNING-FILL MEDIUM < a message
appears on the VFD COMFORT-DISPLAY:

Refill the bath fluid!

2. Switch in stage 2 recognizes a low fluid level

.

If stage 2 of the low level protection according to IEC 61010-2-
010 is triggered, a complete, all-pole shutdown of heater and
circulating pump is effected
A continuous alarm sounds and together with the ticker: > LOW
LEVEL ALARM-FILL MEDIUM
< a message appears on the
VFD COMFORT-DISPLAY:


Turn off the unit with the mains switch, refill bath fluid and turn
the unit on again!

3. Float

4. Circulating pump

5. Heater

Warning:

When adding bath fluid, always us the type of fluid which is identical with the fluid in
the bath.
Bath oils must not contain any water and should be pre-heated approximately to the
current bath temperature! Explosion hazard at high temperatures!