Remote control via the serial interface – JULABO SE-Z Heating Circulators User Manual

Page 39

Advertising
background image

HE / SE

39

Level 2

Level 3

Parameter level

Hours flash, set by pressing

+

Minutes flash, set by pressing

+

Day flashes, set by pressing

+

Month flashes, set by pressing

+

Year flashes, set by pressing

+

Return to factory settings by pressing

RESET returns all set values to the factory setting
except for date and time.

A RESET can be effected only in the –OFF- mode.

During the message –RUN- all parameters are
reset to factory settings.




9.4.1.

Remote control via the serial interface

Factory setting: OFF

The control electronics offer two ways of adjusting a setpoint.

1. Adjustment of setpoint using the keypad

or the integrated programmer.

2. Adjustment of setpoint via the serial interface RS232 using a PC or a

superordinated process control system.

The topline of the VFD-DISPLAY shows a bright „R“ for remote
control; - remote control discontinued.

RS232

IMPORTANT: additional measures for remote control

Connect the circulator to the PC using an interface cable.

Check the interface parameters of both interfaces (circulator and PC)
and make sure they match.
(refer to 12.1.

Setup for remote controll page 71)