Filling / draining – JULABO SE-Z Heating Circulators User Manual

Page 22

Advertising
background image

Preparations

22

5.4.

Filling / draining

Notice:

• Pay attention to the thermal expansion of bath oil during heating to avoid

overflowing of the liquid.

• Do not drain the bath fluid while it is hot!

Recommendation: Temperature range 5 °C to 40 °C
Check the temperature of the bath fluid prior to draining (by switching the unit on
for a short moment, for example).

• Store and dispose the used bath fluid according to the laws for environmental

protection.


Filling

Take care that no liquid enters the interior of the circulator.


• Recommended maximum filling level with water as bath fluid:

30 mm below the tank rim

• Recommended maximum filling level with bath oils:

40 mm below the tank rim

After filling, immerse the samples in the bath or place the lid on
the bath, in case the opening is not to be used.

The circulator provides an early warning system for low level
(description – please refer to page 28) that may be triggered when
changing samples in the bath.

Draining

• Turn the unit off with the mains switch.
• Place the heating circulator near the rim of the table. Use a

suitable vessel as recipient for the bath fluid.

• Unscrew the drain plug on the front of the bath tank to drain the

bath.