Menu program – creation, administration, E 35 – JULABO SE-Z Heating Circulators User Manual

Page 35

Advertising
background image

HE / SE

35

9.2.

MENU PROGRAM – creation, administration

Menu level 1

1 program

10 sections

The integrated programmer permits fast and easy programming of setpoint
temperature sequences. This temperature sequence is called program. A
program is composed of individual sections (STEP). The sections are defined
by duration (TSLICE) and target temperature. The target temperature is the
setpoint (SETPNT), which is achieved at the end of a section. The programmer
calculates the temperature ramp from the difference in time and temperature.

STEP

(Nr.)

1

2

4

6

8

SETPNT (°C)

100

180

180

75

75

TIME

(hh.mm)

00:20

00.10

00:20

00:50

00:20

Graph1

25

50

10 20 30 40 50 60 10 20 30 40 50 60 10 [min]

75

100

150

200

°C

t

Step 1

2

4

6

8


Sections without set value and time are skipped. They can be defined