9.9.1.

Outputs of the connector - REG+E-PROG

CHANNEL 1

CHANNEL 3

CHANNEL 2

REG+E-PROG

EPROG

1.

REG+E-PROG :

1.

Select CHANNEL 1, 2 or 3

2.

First define the desired output value for CHANNELs 1 to 3:

internal actual temperature value (bath temperature))

external actual temperature value (external sensor)

periodic or intermittent heating or cooling

active setpoint temperature

(SETPoint 1, 2, 3,/ integr. programmer /external programmer)

2.

3.

3.

Then select the display size for CHANNELs 1 to 3:

Voltage outputs CHANNELs 1 and 2
Assign the voltage values of 0 V to the lowest and 10 V to the highest
necessary temperature (°C) or power rating (%) required as an output
value.
CHANNEL 3: Output for temperature value (°C) / power (%)
The current output (channel 3) offers 2 ranges for selection:
0 mA to 20 mA or
4 mA to 20 mA
Assign the current values 0 mA or 4 mA to the lowest and 20 mA to
the highest temperature (°C) or power rating (%) required as an
output value.

or

0 V

0 mA

20 mA

4 mA

10 °C

210 °C

197 °C

202 °C

∆ Τ = 5 K

∆ Τ = 200 K

Example 1:

lowest temperature value:

10 °C

highest temperature value

210 °C

Fig. shows 200

°C scaled to paper width

rise: 50 mV/°C

Example 2:

lowest temperature value:

197 °C

highest temperature value: 202 °C

Fig. shows 5

°C scaled to paper width

rise: 2000 mV/°C