3. Instrument Components and Specifications

3.2 Technical data

General
Approvals:

The instrument-specific approval marks are located at the rear panel

of the instrument, next to the name plate.

Nominal supply voltages:

100 to 120 V ~ +/- 10 %

200 to 240 V ~ +/- 10 %

Nominal frequency:

50 to 60 Hz

Power fuse:

2x T 3.15 A L250 V

Maximum power draw at 100 - 120 V:

4.0 A

Maximum power draw at 200 - 240 V:

2.8 A

Leakage current at 240 V/50 Hz:

ca. 2.4 mA

Nominal power:

700 VA

IEC 1010 classification:

Protection class 1, pollution degree 2

Overvoltage installation category II

Operating elevation:

up to max. 2000 m NN

A-weighted noise level:

< 70 dB (A)

IP protection class (IEC 60529)

IP20

Climatic conditions for the operation of the instrument:

Temperature:

+15 °C to +30 °C

Relative humidity:

20 - 80 % - non-condensing

Climatic conditions for the storage and transport of the packaged instrument:

Storage temperature range:

+5 °C to +40 °C

Transport temperature range:

-29 °C to +40 °C

Relative humidity:

10 - 85 % - non-condensing

Dimensions and weight
Dimensions of basic instrument

Width x depth:

475 x 650 mm

Height with magazine:

895 mm

Dimensions with unload station connected

Width x depth:

548 x 650 mm

Height with magazine:

655 mm

Basic instrument empty weight:

approx. 28 kg

Weight, packed:

approx. 65 kg

Unload station empty weight:

approx. 14 kg

Weight, packed:

approx. 32 kg