Troubleshooting, Error messages – Leica Biosystems IPC User Manual

Page 65

Advertising
background image

63

Leica IP C

7. Troubleshooting

Error messages

(continued)

Display

code

Source of error

Troubleshooting solution

Cover (

4 in Fig. 2

) not closed properly.

Safety switch triggered.

Ensure that the cover is not blocked, e.g. by the

manual unload station. Close cover completely.

Flash standby state not reached within the pre-

scribed time. Charging electronics defective.

Call Leica Technical Service.

No flash or flash duration too short.

Flash bulb dirty or defective.

Check whether flash bulb works – to do so,

observe the stray light on the cover.

Never open the cover to check whether a

flash is triggered!

No flash --> install a new flash bulb. See

Chapter 4.6

.

The temperature in the drying module is not

within the permissible temperature range.

Carry out a RESET. See

Chapter 7.1

.

Unload station mechanically blocked.

Remove the cause of the blockage.

Heater fan not running or running too slowly.

Call Leica Technical Service.

No cassette is in the drying module for flashing.

Cassette left the cassette carrier, but did not

reach the drying module.

Remove cassette.

Cassette not successfully ejected from drying

module.

Remove cassette.

Cassette in the drying module during initialization

or prior to a print job being carried out.

Remove the cause of the blockage.

The flap of the drying module either does not

close or does not open. The flap is blocked (e.g.

by a cassette).

Remove the cause of the blockage.

Lifting table end sensor does not switch.

Call Leica Technical Service.

Lifting table position sensor does not switch dur-

ing positioning.

Call Leica Technical Service.

Flash power supply: Overtemperature for more

than 10 minutes.

Call Leica Technical Service.