Running a test print

• Run a test print to verify whether the print head works correctly.
• For that purpose, fill some cassettes into a magazine and insert the mag-

azine into magazine position 1.

• Press and hold the CLEAN button until "00" is displayed, then release the

button. A cassette is imprinted with a stored test image. If the print result

is not satisfactory, this step can be repeated several times.

At this point, the instrument has to be "told" which type of cartridge has been

inserted.

There are three options:
1. New ink cartridge:

Press

LOADED; the printer sets the ink level meter to 'full'.

2. Used ink cartridge:

Press

ERROR; the printer resumes measuring at the ink level where it

previously left off.

• The sensor in the cartridge slot recognizes the presence of a

new cartridge.

• The "Ink Empty" indicator LED goes out and "88" appears on the

display.

4. Instrument Setup

Exchanging the cartridge

(continued)

р

р

CAUTION!

NEVER press CLEAN while an ink cartridge is in the instrument! The

entire contents of the ink cartridge will spill into the printer.

The fill level of the cleaning cartridge is not monitored. Each use

should be noted on the cartridge. The cartridge can be used twice.

The cycle time when inserting a cleaning cartridge is 3.5 minutes and

is thus considerably longer than that of an ink cartridge.

• After one of the three buttons has been pressed, the ink exchange software

routine starts; air is evacuated from the hoses and the system is refilled

with liquid.

• When the procedure has been completed, "88" disappears from the display.

3. Used or new storage fluid cartridge:

Press the key CLEAN; the current ink level is stored.