Troubleshooting, 1 malfunctions – Leica Biosystems IPC User Manual

Page 62

Advertising
background image

60

Instructions for Use V 1.9 RevD – 05/2013

7. Troubleshooting

How to eliminate the problem:
• Determine the cause of the error using the error list (

Chapter 7.3)

.

• Remove obstruction(s); open lid if required.
• Important!

Remove all cassettes which are still in the chute, in or next to the

cassette carrier or in the drying module!

These cassettes should not be reused.
• Close the lid and press ERROR to confirm to the printer that the source

of the error has been eliminated.

• The printer then verifies whether all processing paths are unobstructed

and whether the source of the error has been eliminated.

• If there are still some obstructions left, or if the source of error has not

been thoroughly eliminated, the printer displays another error message.

• Interrupted print jobs are resumed where previously left off.
• If an error message is displayed several times although all possible

causes have been eliminated, a Reset should be carried out.

Reset:

• Press and release LOADED and ERROR simultaneously.
• A Reset restores the printer to the same state as directly after switching

on. All print jobs in the print queue are deleted.

• If the same error continues to be displayed even after a reset, switch the

printer off via the power switch (back panel) and, after a short waiting

period of approx. 30 sec., switch it back on. If this does not eliminate the

problem either, call Leica Technical Service.

+

+

Confirm elimination

of error

7.1 Malfunctions

If the printer malfunctions during printing, a corresponding error

code is displayed and simultaneously the LED in the ERROR button

starts flashing.

Advertising