Leica IP C

• Push a (new) cleaning cartridge as far as it

will go into the slot (

Fig. 62

).

• Unscrew the red cap (85) one turn and slide

the cartridge fully home.

• Push the security latch (86, in

Fig. 61

) back in

front of the cartridge.

+

Instructions for storage

(continued)

Fig. 63

72

69

Fig. 62

• Push the small lever (69,

Fig. 63)

upwards

under the print head.

64

85

• Open the hood of the printer

(4 in

Fig. 36

), then press the

CLEAN and LOADED keys simul-

taneously.

• The print head (72) moves up-

ward to a position approx. 1 cm

( inch) away from the sealing

lip (

Fig. 63

).

• The "Ink Empty" LED will go out and "88" will

appear on the display.

• Press CLEAN to rinse the print head; "00" will

appear on the display. The display will go out

when the cleaning process is complete.

6. Cleaning and Maintenance

р

р

The cleaning process with solvent

takes approx. 3.5 minutes.