Cleaning and maintenance – Leica Biosystems IPC User Manual

Page 61

Advertising
background image

59

Leica IP C

Transporting the printer

If the printer is to be shipped, the storage instructions described above

are to be carried out completely. Then, additionally make the following

preparations:

• Open the hood (4 in

Fig. 36)

of the printer and loosen the screw of the

flash cover.

• Then take out the flash bulb. For more information, refer to

Chap. 4.6

of

this manual

• Close the flash cover and the hood.

• Use original packaging and securely screw the printer to the baseplate

(see unpacking instructions).

• Reinsert the transport anchor (27) (see

Fig. 16

) and secure the hood with

adhesive tape.

• Make sure that the instrument is only transported upright.

• Likewise, clean the removed location plate (70)

with the sealing lip with (clean) alcohol and a

cleaning swab (89) (

Fig. 66

).

• The sealing lip (71) must be completely free of

ink residue. Check the sealing lip for damage.

Do not reuse a replacement plate with a dam-

aged sealing lip.

• Clean the instrument completely, as described

in this chapter.

Instructions for storage

(continued)

Fig. 66

70

71

89

6. Cleaning and Maintenance