Instrument setup – Leica Biosystems IPC User Manual

Page 31

Advertising
background image

29

Leica IP C

Fig. 24

45

47

Inserting the new flash bulb

47

48

Fig. 26

Fig. 23

Fig. 25

45

44

4. Instrument Setup

44

46

85

86

86

• First insert a shielding glass (85) into the two

holders (86) (

Fig. 23

).

• Insert the new flash bulb into socket (46,

Fig. 25

); then push it carefully inwards as far

as it will go (

Fig. 26

) (the polarity marker (+)

must not be visible any longer). If necessary,

move flash bulb gently up and down.

• Make sure that the bulb is inserted correctly.

The electrode marked with + (47) has to be

inserted into the socket (46) with the same

mark (48,

Fig. 25

).

Caution!

If the bulb electrodes are inserted the

wrong way, the flash bulb will still

function, but the bulbs life will be con-

siderably reduced.

• The contact spring (44) has to touch the igni-

tion wire (45) of the bulb after being inserted.

• Swing the reflector downwards. Reinsert and

retighten screw (49).

• Close lid (4) of the instrument again.