6 installing/exchanging the flash bulb, Installing/exchanging the flash bulb, Instrument setup – Leica Biosystems IPC User Manual

Page 30

Advertising
background image

28

Instructions for Use V 1.9 RevD – 05/2013

Removing the old flash bulb

4.6 Installing/exchanging the flash bulb

49

41

Fig. 19

• Swing the reflector (41) upwards.

• Carefully pull out the old flash bulb (43) straight

to the right, do not twist it. If the flash bulb

cannot be pulled out easily, gently rock it back

and forth to get it unseated from the socket.

• Make sure the contact spring (44) is removed

from the priming wire (45) of the bulb.

(See also

Fig. 18

and

21

)

Fig. 20

Fig. 21

• Open the lid (4,

Fig. 6

) to gain access to the

reflector.

• Remove screw (49) (use screwdriver supplied

as part of tool set). Mind the washer (42).

Fig. 22

42

43

44

4. Instrument Setup

Switch the instrument off and unplug

it from power supply. Allow the flash

bulb to cool off before removing it. Do

not handle the flash bulb with bare

hands. Use a glove or tissue.

Caution!

To insert/remove, hold the flash

bulb as shown in

Fig. 21

(right).

Do NOT take hold of the sides of

the flash bulb or compress it as

shown in

Fig. 20

(risk of break-

age).