7.6 Replacing the secondary fuses

Replacing the fuses

91

90

• Insert a screwdriver (91) into the slot in fuse

holder (90); push slightly inwards and at the

same time rotate the screwdriver a

1

/

4

turn to

the left.

• The fuse holder is released and can be

removed.

90

62

• Remove the defective fuse (92) from the fuse

holder (90) and insert the correct type replace-

ment fuse.

• Insert the fuse holder with the replacement

fuse. Push the holder in with the screwdriver

and secure it with a

1

/

4

turn to the right.

Fig. 69

Fig. 70

7. Troubleshooting

Before replacing a fuse, always switch the instrument off and unplug from power supply.

Defective fuses may be replaced only with the replacement fuses supplied together with the

instrument.