Cleaning and maintenance, 1 cleaning the instrument, Cleaning the instrument – Leica Biosystems IPC User Manual

Page 54

Advertising
background image

52

Instructions for Use V 1.9 RevD – 05/2013

6. Cleaning and Maintenance

Cleaning of the following IP modules signed by

an arrow is of particular importance:

Cassette guiding mechanisms

• Loading station (

Fig. 54

)

The ejection unit of the magazines, magazine

holders and chute.

Always ensure that the sensor at the end of

the chute is clean.

• Transport station (

Fig. 56

)

Remove dust and debris from the Cassette-

Clamp.

Fig. 54

Fig. 55

6.1 Cleaning the instrument

Prior to cleaning the instrument, always switch off power supply and unplug the power cord!

When handling cleaning detergents, follow the instructions of the manufacturer and make sure

all laboratory regulations in force in your country are complied with.

To clean the exterior surfaces, use a mild and ph-neutral commercial household cleaner. You

may NOT use: Alcohol, cleaning materials containing alcohol (glass cleaner!), abrasives, or

solvents containing acetone or xylene!

No liquid may come into contact with the electrical connections or spill into the interior of the

instrument!

The Leica IP C needs to be thoroughly vacuum cleaned once a week.