4 changing the flash bulb, 5 power failure, Changing the flash bulb – Leica Biosystems IPC User Manual

Page 67: Power failure, Troubleshooting

Advertising
background image

65

Leica IP C

Malfunction / failure
• Instrument not functioning.

• No display indication.
• Instrument not working at normal speed.

Printing a cassette takes approx. 8 sec.,

even after the warm-up phase has been

completed.

• Check whether there is a general power failure (no power).
• Check whether the power plug is inserted correctly into the wall outlet

and whether the wall outlet is switched on, if applicable.

• Check whether the power switch is switched on correctly. The primary

fuse may have tripped. If so, the main switch will be in "0" = OFF position

(

Fig. 67

).

• Check whether one of the two secondary fuses is defective (F1, F2 on

the back panel of instrument,

Fig. 68, see also Fig. 69

).

Some instrument malfunctions / failures are caused by defective fuses.

Fuse F1

Fuse F2

7.5 Power failure

Fuse to be checked

7.4 Changing the flash bulb

Code "13" appears on the printer display when the flash bulb has reached

its maximum service life.
When code 13 is issued, the bulb has to be changed.
For details on how to insert / replace the flash bulb, please refer to

chapter 4.6

.

10 s

Fig. 68

Fig. 67

7. Troubleshooting

ð

If a defective flash bulb has been replaced without the message

"13" appearing on the display, likewise confirm this using the above

procedure (hold down the ERROR key for at least 10 sec.).

Before replacing the flash bulb, switch off the printer and unplug

the instrument from power supply.

After changing the bulb, switch the printer back on.
Then go to offline mode and hold down the ERROR key for 10 seconds; code

"13" disappears from the display.