48

Instructions for Use V 1.9 RevD – 05/2013

5.3 Alarm functions

The Leica IP C is equipped with two different alarm functions:

Instrument alarm

The printer has a beeper that emits acoustic signals indicating important

instrument states and functions.
• Upon pressing the button:

1 short beeping sound

• Magazine empty/tray stack full:

2 short beeping sounds

• In case of error:

5 short beeping sounds

• When ending the head cleaning:

5 short beeping sounds

The beeper can be deactivated

by means of the DIL switches

at the back panel of the printer.
To deactivate the beeper, push

the switch at the very bottom

(60) to the right

(in the direction of the red

arrow in

Fig. 50)

.

Fig. 50

60

Remote alarm

Additionally, an alarm can be installed outside the room in which the printer

works.
• The remote alarm device is connected to the printer via a 3.5 mm-dia-

meter jack connector (optional accessory) that is inserted into socket

(56).

• The remote alarm is triggered if no power is supplied to the printer or if

the power switch at the back panel of the printer is switched off.

56

Back panel of instrument

5. Operation

The remote alarm device connected to the instrument must be rated

at less than 100 mA. A maximum voltage of 24 V DC must not be ex-

ceeded.

For details on how to connect a remote alarm device to the Leica

Printer, please contact your local Leica sales office or the man-

ufacturer directly.