Display

code

Source of error

Troubleshooting solution

Magazine output mechanically blocked.

Remove the cause of the blockage.

Faulty ejection of a cassette. Magazine output blocked. ‘Remove cassette.

Transfer of a cassette from the chute to the cassette

carrier failed. Horizontal motor either incorrectly posi-

tioned or mechanically blocked.

Remove cassette.

Cassette is stuck in the feeding chute.

Remove cassette.

Horizontal drive is mechanically blocked.

Remove cassette.

Vertical drive is mechanically blocked.

Remove cassette.

Rotational motion is mechanically blocked.

Remove cassette.

Cassette not correctly clamped in cassette carrier.

Cassette did leave the feeding chute but did not reach

the cassette carrier.

Remove the cassette from the cassette

carrier.

Cassette did not leave the cassette carrier or was still

located in the cassette carrier during initialization.

Remove the cassette from the cassette

carrier.

The ink print head is getting too hot.

Ambient temperature too high or electronics defective

Switch instrument off and let it cool off.

The instrument remains disabled until

the print head has cooled down to a

temperature value within the permis-

sible range.

Check ambient temperature.

No or incorrect voltage at ink print head.

Call Leica Technical Service.