5. Set Fixture Displacement Limits.

a) Calculate Fixture Displacement Limits.

When the actuator is positioned to touch the uncompressed spring, the end of the actuator will be
90 mm above the table. Since full stroke is only 70 mm, no limit restriction is required. Fixture limits
should be set at +35 mm and -35 mm.

b) Set the Fixture Displacement Limits to ±35 mm.

110 | MTS Acumen

Running the Example Spring Test