background image

Click the green arrow button to create the compensation set. The stages of verification are shown
above the compensation progress bar. When verification is complete, the bar disappears.

The mass of the fixturing and load cell is calculated, and the results are indicated by the arrow on the
Correction Quality Indicator. Optimal results are achieved when the arrow appears in the green area.

The accelerometer used to measure the inertial force that requires offset is configured to output a value
of 1 g when the actuator is not moving. This 1 g signal creates a static offset that is dependent on the
mass of the grips and fixtures attached to the actuator. This static offset cannot be cleared.

5. If necessary, adjust the compensation set.

If you are not satisfied with the results:

a) Make adjustments to the parameters or move the slider.
b) Click Adjust.
c) Rerun the compensation command.

6. Save the compensation set.

120 | MTS Acumen

Running the Example Spring Test