MTS Acumen User Manual

Page 148

Advertising
background image

Your system is equipped with identical detectors in both the controller (set with the Station Manager
application) and in MPX. To ensure detectors do not conflict with each other and cause unwanted
shutdowns, MTS configures MTS Acumen Station Configurations to enable detectors in MPX and
disable detectors in the controller (with the exception of the Axial Acceleration Limit). This step ensures
all controller settings are set properly.

a) Click the Detectors icon in the Station Manager main window.

b) With All Detectors selected in the List box, ensure that all detectors under the Error Detectors

tab are disabled.

c) With All Detectors selected in the List box, ensure that all detectors under the Limit Detectors

tab except Axial Acceleration are disabled.

d) Ensure that the settings for Axial Acceleration are as shown:

148 | MTS Acumen

Best Practices for Other System Configurations and Tests

Advertising