Limited warranty, Warning – SATEC PM180 Operation Manual User Manual

Page 3

Advertising
background image


LIMITED WARRANTY

The manufacturer offers the customer a 24-month functional warranty on the
instrument for faulty workmanship or parts from date of dispatch from the distributor.
In all cases, this warranty is valid for 36 months from the date of production. This
warranty is on a return to factory basis.
The manufacturer does not accept liability for any damage caused by instrument
malfunction. The manufacturer accepts no responsibility for the suitability of the
instrument to the application for which it was purchased.
Failure to install, set up or operate the instrument according to the instructions herein
will void the warranty.
Only a duly authorized representative of the manufacturer may open your instrument.
The unit should only be opened in a fully anti-static environment. Failure to do so
may damage the electronic components and will void the warranty.
The greatest care has been taken to manufacture and calibrate your instrument.
However, these instructions do not cover all possible contingencies that may arise
during installation, operation or maintenance, and all details and variations of this
equipment are not covered by these instructions.
For additional information regarding installation, operation or maintenance of this
instrument, contact the manufacturer or your local representative or distributor.

WARNING

Read the instructions in this manual before performing installation, and take note
of the following precautions:

Ensure that all incoming AC power and other power sources are turned OFF
before performing any work on the instrument. Failure to do so may result in
serious or even fatal injury and/or equipment damage.

Before connecting the instrument to the power source, check the labels on
the back of the instrument to ensure that your instrument is equipped with
the appropriate power supply voltage, input voltages and currents.

Under no circumstances should the instrument be connected to a

power source if it is damaged.

To prevent potential fire or shock hazard, do not expose the instrument
to rain or moisture.

The secondary of an external current transformer must never be allowed to
be open circuit when the primary is energized. An open circuit can cause
high voltages, possibly resulting in equipment damage, fire and even
serious or fatal injury. Ensure that the current transformer wiring is secured
using an external strain relief to reduce mechanical strain on the screw

terminals, if necessary.

Only qualified personnel familiar with the instrument and its associated
electrical equipment must perform setup procedures.

Do not open the instrument under any circumstances when it is connected
to a power source.

Do not use the instrument for primary protection functions where failure of
the device can cause fire, injury or death. The instrument can only be used
for secondary protection if needed.

Read this manual thoroughly before connecting the device to the current carrying

circuits. During operation of the device, hazardous voltages are present on input
terminals. Failure to observe precautions can result in serious or even fatal injury or
damage to equipment.


All trademarks are property of their respective owners.

October 2012

Copyright 2003-2012

SATEC Ltd.

Advertising